UNIFORM DEPTH DOSE DISTRIBUTION FOR BIOLOGICAL IRRADIATION USING NEGATIVE PIONS

被引:6
作者
LAM, GKY [1 ]
HENKELMAN, RM [1 ]
HARRISON, RW [1 ]
SKARSGARD, LD [1 ]
PALCIC, B [1 ]
机构
[1] UNIV BRITISH COLUMBIA,FAC BATHO BIOMED,TRIUMF,VANCOUVER V6T 1W5,BC,CANADA
关键词
D O I
10.1088/0031-9155/24/6/014
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A simple, flexible technique has been developed to generate uniform depth dose profiles for the biomedical pion beam at TRIUMF using dynamic momentum control and linear programming. Either the entrance dose or the irradiation time required for a certain dose over the uniform region can be minimised. The dynamic momentum control can operate automatically under computer control even with a highly unstable beam. Cell survival profiles have been obtained for this uniform dose distribution using the gelatin technique. The RBE increases with increasing depth through the uniform dose region.
引用
收藏
页码:1243 / 1249
页数:7
相关论文
共 14 条
[1]  
Dantzig G., 1951, ACTIVITY ANAL PRODUC, P339
[2]  
DICELLO JF, 1975, P PARTICLE RAD THERA
[3]  
GASS SI, 1969, LINEAR PROGRAMMING M
[4]   NEGATIVE PION BEAM TRANSPORT CHANNEL FOR RADIOBIOLOGY AND RADIATION-THERAPY AT TRIUMF [J].
HARRISON, RW ;
LOBB, DE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1973, NS20 (03) :1029-1031
[5]  
HENKELMAN RM, 1977, INT J RADIAT ONCOL, V2, P123, DOI 10.1016/0360-3016(77)90017-7
[6]  
HENKELMAN RM, 1978, 6TH S MICR BRUSS, V1, P497
[7]  
HENKELMAN RM, 1977, TRI772 TRIUMF EXT RE
[8]   RANGE MODULATORS FOR PROTONS AND HEAVY-IONS [J].
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM .
NUCLEAR INSTRUMENTS & METHODS, 1975, 131 (03) :437-440
[9]   AUTOMATED DOSE MAPPING SYSTEM FOR TRIUMF BIOMEDICAL PION BEAM [J].
LAM, GKY ;
HENKELMAN, RM ;
HARRISON, RW .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (04) :768-776