In TV-based megavoltage imaging systems it is useful to be able to estimate the geometrical efficiency g2 of the camera lens. It is shown that the appropriate expression is g2 = (16n2)-1 x [F(1 + 1/m)]-2, where n is the refractive index of the scintillator, F is the F-number of the lens and m is the optical magnification. This expression yields estimates for g2 that are 5 to 10 times smaller than the estimates that have appeared previously in the literature.
引用
收藏
页码:1152 / 1153
页数:2
相关论文
共 6 条
[1]  
BARRETT HH, 1981, RADIOLOGICAL IMAGING, P288
[2]   LIGHT ANGULAR-DISTRIBUTION OF FLUORESCENT SCREENS EXCITED BY X-RAYS [J].
GIAKOUMAKIS, GE ;
MILIOTIS, DM .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (01) :21-29
[3]   A DIGITAL FLUOROSCOPIC IMAGING DEVICE FOR RADIOTHERAPY LOCALIZATION [J].
MUNRO, P ;
RAWLINSON, JA ;
FENSTER, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (03) :641-649
[4]   AN EVALUATION OF A FLUORESCENT SCREEN-ISOCON CAMERA SYSTEM FOR X-RAY-IMAGING IN RADIOLOGY [J].
NELSON, RS ;
BARBARIC, ZL ;
GOMES, AS ;
MOLER, CL ;
DECKARD, ME .
MEDICAL PHYSICS, 1982, 9 (05) :777-783
[5]   VIDEO TECHNIQUES FOR ONLINE PORTAL IMAGING [J].
SHALEV, S ;
LEE, T ;
LESZCZYNSKI, K ;
COSBY, S ;
CHU, T .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1989, 13 (03) :217-226
[6]  
1974, EOH11 RCA CORP EL HD, P216