OPTICAL DESIGN AND PERFORMANCE OF THE INELASTIC-SCATTERING BEAMLINE AT THE NATIONAL SYNCHROTRON LIGHT-SOURCE

被引:45
作者
KAO, CC
HAMALAINEN, K
KRISCH, M
SIDDONS, DP
OVERSLUIZEN, T
HASTINGS, JB
机构
[1] UNIV HELSINKI,DEPT PHYS,SF-00014 HELSINKI,FINLAND
[2] EUROPEAN SYNCHROTRON RADIAT FACIL,F-38043 GRENOBLE,FRANCE
关键词
D O I
10.1063/1.1146467
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Phase I of the X21 beamline at the National Synchrotron Light Source was commissioned during 1993. The research program at the X21 beamline is focused on the study of electronic excitations in condensed matter with total energy resolution of 0.1-1.0 eV. The source is a 27-pole hybrid wiggler. A water-cooled horizontal focusing Si(220) monochromator and a spherically bent Si(444) analyzer were installed and commissioned. At 8 keV the energy resolution of the monochromator is about 0.7 eV, and the energy resolution of the analyzer is about 0.1 eV. Results from several selected experiments are also discussed. © 1995 American Institute of Physics.
引用
收藏
页码:1699 / 1702
页数:4
相关论文
共 13 条
  • [1] OPTICAL DESIGN AND PERFORMANCE OF THE X25 HYBRID WIGGLER BEAM LINE AT THE NATIONAL SYNCHROTRON LIGHT-SOURCE
    BERMAN, LE
    HASTINGS, JB
    OVERSLUIZEN, T
    WOODLE, M
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) : 428 - 432
  • [2] MAGNETIC MEASUREMENTS ON INSERTION DEVICES AT NSLS
    DECKER, G
    GALAYDA, J
    SOLOMON, L
    KITAMURA, M
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) : 1845 - 1848
  • [3] HAMALAINEN K, UNPUB
  • [4] KRISCH MH, UNPUB
  • [5] X-RAY CURVED-CRYSTAL MONOCHROMATOR SYSTEM AT STORAGE RING DCI
    LEMONNIER, M
    FOURME, R
    ROUSSEAUX, F
    KAHN, R
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1978, 152 (01): : 173 - 177
  • [6] X-RAY COMPTON-SCATTERING STUDY OF ELECTRONIC DISTRIBUTION IN BERYLLIUM
    LOUPIAS, G
    PETIAU, J
    ISSOLAH, A
    SCHNEIDER, M
    [J]. PHYSICA STATUS SOLIDI B-BASIC RESEARCH, 1980, 102 (01): : 79 - 95
  • [7] AN X-RAY SPECTROMETER FOR COMPTON-SCATTERING
    SAKURAI, Y
    ITO, M
    URAI, T
    TANAKA, Y
    SAKAI, N
    IWAZUMI, T
    KAWATA, H
    ANDO, M
    SHIOTANI, N
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) : 1190 - 1193
  • [8] INTERBAND-TRANSITIONS AND CORE EXCITATION IN HIGHLY ORIENTED PYROLYTIC-GRAPHITE STUDIED BY INELASTIC SYNCHROTRON X-RAY-SCATTERING - BAND-STRUCTURE INFORMATION
    SCHULKE, W
    BONSE, U
    NAGASAWA, H
    KAPROLAT, A
    BERTHOLD, A
    [J]. PHYSICAL REVIEW B, 1988, 38 (03): : 2112 - 2123
  • [9] SCHULKE W, 1991, HDB SYNCHROTRON RAD, V3, pCH15
  • [10] AN X-RAY SPECTROMETER FOR COMPTON-SCATTERING EXPERIMENTS WITH SYNCHROTRON RADIATION
    SHIOTANI, N
    SAKAI, N
    ITOH, F
    SAKURAI, M
    KAWATA, H
    AMEMIYA, Y
    ANDO, M
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 275 (02) : 447 - 452