ENERGY-CONSERVING PROGRAMMING OF VVI PACEMAKERS - A TELEMETRY-SUPPORTED, LONG-TERM, FOLLOW-UP-STUDY

被引:6
作者
KLEIN, HH
KNAKE, W
机构
[1] UNIV GOTTINGEN,DEPT CARDIOL,W-3400 GOTTINGEN,GERMANY
[2] UNIV GOTTINGEN,DEPT CARDIOTHORAC SURG,W-3400 GOTTINGEN,GERMANY
关键词
battery longevity; programming; VVI pacemaker;
D O I
10.1002/clc.4960130608
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Thirty patients with VVI pacemakers (Quantum 253‐09, 253‐19, Intermedics Inc., Freeport, TX) were observed for a mean of 65 months. Within 12 months after implantation, optimized output programming was performed in 29 patients. This included a decrease in pulse amplitude (22 patients), pulse width (4 patients), and/or pacing rate (11 patients). After 65 months postimplantation, telemetered battery voltage and battery impedance were compared with the predicted values expected when the pulse generator constantly stimulates at nominal program conditions (heart rate 72.3 beats/min, pulse amplitude 5.4 V, pulse width 0.61 ms). Instead of an expected cell voltage of 2.6 V and a cell impedance of 10 kΩ mean telemetered values amounted to 2.78 V and 1.4 kΩ, respectively. These data correspond to a battery age of 12‐15 months at nominal program conditions. This longterm follow‐up study suggests that adequate programming will extend battery longevity and thus pulse generator survival in many patients. Copyright © 1990 Wiley Periodicals, Inc.
引用
收藏
页码:409 / 413
页数:5
相关论文
共 11 条
[1]  
ASTRINSKY EA, 1983, CLIN PROG PACING ELE, V1, P51
[2]   SUCCESSFUL MANAGEMENT OF PACING SYSTEM MALFUNCTIONS WITHOUT SURGERY - THE ROLE OF PROGRAMMABLE PULSE GENERATORS [J].
BILLHARDT, RA ;
ROSENBUSH, SW ;
HAUSER, RG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (05) :675-682
[3]   TRANSTELEPHONIC MONITORING OF 25,919 IMPLANTED PACEMAKERS [J].
DREIFUS, LS ;
ZINBERG, A ;
HURZELER, P ;
PUZIAK, AD ;
PENNOCK, R ;
FELDMAN, M ;
MORSE, DP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (03) :371-378
[4]   SOLID-STATE LITHIUM BATTERY - NEW IMPROVED CHEMICAL POWER SOURCE FOR IMPLANTABLE CARDIAC PACEMAKERS [J].
GREATBATCH, W ;
LEE, JH ;
MATHIAS, W ;
ELDRIDGE, M ;
MOSER, JR ;
SCHNEIDER, AA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1971, BM18 (05) :317-+
[5]   PACEMAKER FOLLOW-UP - ITS ROLE IN THE DETECTION AND CORRECTION OF PACEMAKER SYSTEM MALFUNCTION [J].
GRIFFIN, JC ;
SCHUENEMEYER, TD ;
HESS, KR ;
GLAESER, D ;
ANDERSON, BJ ;
ROMANS, E ;
JENKINS, MA ;
NIELSEN, AP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (03) :387-391
[6]   12 YEARS OF CLINICAL-EXPERIENCE WITH LITHIUM PULSE GENERATORS [J].
HAUSER, RG ;
WIMER, EA ;
TIMMIS, GC ;
GORDON, S ;
STALLER, B ;
KLODNYCKY, M ;
EISENHAUER, A ;
GOREN, C ;
MORAN, JF ;
DAVIS, AA ;
WALSH, R ;
MCDONOUGH, T ;
LAWLESS, C ;
WYNDHAM, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06) :1277-1281
[7]   PROGRAMMING PACEMAKERS - A GAIN OR A GIMMICK [J].
IRNICH, W .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1985, 110 (40) :1527-1530
[8]   THE CHRONAXIE TIME AND ITS PRACTICAL IMPORTANCE [J].
IRNICH, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1980, 3 (03) :292-301
[9]  
Klein H H, 1987, Pacing Clin Electrophysiol, V10, P152, DOI 10.1111/j.1540-8159.1987.tb05938.x
[10]   EXPERIENCE WITH TELEMETRY-SUPPORTED PACEMAKER CONTROL IN PATIENTS WITH VVI-PACEMAKERS [J].
KLEIN, HH ;
HANH, BK ;
HELLBERG, K ;
RUSCHEWSKI, W ;
DEVIVIE, ER ;
KREUZER, H .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1985, 110 (38) :1447-1451