THE IMPACT OF AUTOMATED BLOOD-PRESSURE DEVICES ON THE EFFICIENCY OF CLINICAL-TRIALS

被引:10
作者
APPEL, LJ [1 ]
MARWAHA, S [1 ]
WHELTON, PK [1 ]
PATEL, M [1 ]
机构
[1] JOHNS HOPKINS HLTH INST,WELCH CTR PREVENT EPIDEMIOL & CLIN RES,BALTIMORE,MD 21205
来源
CONTROLLED CLINICAL TRIALS | 1992年 / 13卷 / 03期
关键词
BLOOD PRESSURE DETERMINATION; BLOOD PRESSURE MONITORING; COMPARATIVE STUDY; HYPERTENSION;
D O I
10.1016/0197-2456(92)90006-L
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
By reducing measurement error, automated blood pressure (BP) devices should enhance the precision of BP estimation and thereby decrease sample size requirements in clinical trials of BP-lowering therapy. Enhanced precision would be particularly relevant to clinical trials assessing the efficacy of nonpharmacological therapies. In the present investigation, resting (clinic) BPs by the Dinamap 8100 (a stationary device) and the Accutracker II (an ambulatory device) were as precise as manual BPs given an equal number of observations by each method. However, both the Dinamap and Accutracker devices underestimated resting diastolic BP in comparison to the manual observers. Estimates of average daytime and 24-hour ambulatory BP, based on large numbers of observations over an extended period of time, were extremely precise. These findings suggest that the use of automated devices to measure resting BP may not reduce samples sizes, whereas use of ambulatory BP devices should reduce samples sizes considerably.
引用
收藏
页码:240 / 247
页数:8
相关论文
共 16 条
[1]   THE ACCURACY AND PRECISION OF THE ACCUTRACKER AMBULATORY BLOOD-PRESSURE MONITOR [J].
APPEL, LJ ;
WHELTON, PK ;
SEIDLER, AJ ;
PATEL, AR ;
KLAG, MJ .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 1990, 132 (02) :343-354
[2]  
APPEL LJ, 1990, 13TH P SCI M INT SOC, pS88
[3]  
CONWAY J, 1988, J HYPERTENS, V6, P111
[4]  
CONWAY J, 1989, J HYPERTENS, V7, pS29
[5]  
CURB JD, 1983, HYPERTENSION, V4, P610
[6]  
HARSHFIELD GA, 1989, HDB RES METHODS CARD, P193
[7]  
Hypertension Prevention Trial Research Group, 1990, ARCH INTERN MED, V150, P153, DOI [10.1001/archinte.1990.00390130131021, DOI 10.1001/ARCHINTE.1990.00390130131021]
[8]   INTRODUCTION TO SAMPLE-SIZE DETERMINATION AND POWER ANALYSIS FOR CLINICAL-TRIALS [J].
LACHIN, JM .
CONTROLLED CLINICAL TRIALS, 1981, 2 (02) :93-113
[9]   BLOOD-PRESSURE DURING NORMAL DAILY ACTIVITIES, SLEEP, AND EXERCISE - COMPARISON OF VALUES IN NORMAL AND HYPERTENSIVE SUBJECTS [J].
PICKERING, TG ;
HARSHFIELD, GA ;
KLEINERT, HD ;
BLANK, S ;
LARAGH, JH .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1982, 247 (07) :992-996
[10]   PREDICTIVE VALUES OF ROUTINE BLOOD-PRESSURE MEASUREMENTS IN SCREENING FOR HYPERTENSION [J].
ROSNER, B ;
POLK, BF .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 1983, 117 (04) :429-442