INTERNAL CONVERSION ELECTRON MEASUREMENT BY A SI SURFACE BARRIER DETECTOR

被引:4
作者
TAMURA, T
KUROYANAGI, T
机构
[1] Division of Physics, Tokai Research Establishment, Japan Atomic Energy Research Institute, Ibaraki-ken, Tokai-mura
来源
NUCLEAR INSTRUMENTS & METHODS | 1969年 / 67卷 / 01期
关键词
D O I
10.1016/0029-554X(69)90538-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
It is shown that combination of a Si surface barrier detector and a paralleled FETs preamplifier under the cooled condition gives a resolution of 2.6-keV fwhm for the K-conversion electron of the 136-keV transition in 57Fe. This method can be applied with many advantages for the measurements of low energy conversion electrons from short-lived activities. © 1969.
引用
收藏
页码:38 / +
页数:1
相关论文
共 5 条
[1]  
BLANKENSHIP JL, 1960, IRE T NUCL SCI, VNS7, P190
[2]  
FOX RJ, 1962, IRE T NUCL SCI, VNS9, P213
[3]  
HANSEN NJ, 1962, IRE T NUCL SCI, VNS9, P217
[4]  
KUROYANAGI T, TO BE PUBLISHED
[5]   ACCURATE DETERMINATION OF IONIZATION ENERGY IN SEMICONDUCTOR DETECTORS [J].
PEHL, RH ;
GOULDING, FS ;
LANDIS, DA ;
LENZLING.M .
NUCLEAR INSTRUMENTS & METHODS, 1968, 59 (01) :45-&