ARGON INDIRECT OPHTHALMOSCOPIC PHOTOCOAGULATION - REDUCED POTENTIAL PHOTOTOXICITY WITH A FIXED SAFETY FILTER

被引:6
作者
WHITACRE, MM [1 ]
MANOUKIAN, N [1 ]
MAINSTER, MA [1 ]
机构
[1] COHERENT MED DIV,PALO ALTO,CA
关键词
D O I
10.1136/bjo.74.4.233
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A new argon indirect ophthalmoscopic photocoagulator is presented which uses a red helium-neon laser aiming beam. An interference filter protects the operator from green or blue-green treatment beam reflections without impairing visualisation of the aiming beam or significant retinal anatomy. The protective filter is fixed in place, eliminating the weight, noise, and potential failure of mechanically switched filters. The red aiming beam has a negligible potential for producing photochemical retinal damage in the patient or operator.
引用
收藏
页码:233 / 234
页数:2
相关论文
共 13 条
[1]   CLINICAL-EXPERIENCE WITH A BINOCULAR INDIRECT OPHTHALMOSCOPE LASER DELIVERY SYSTEM [J].
FRIBERG, TR .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 1987, 7 (01) :28-31
[2]  
GREISS GA, 1981, INVEST OPHTHALMOL VI, V20, P803
[3]   AN AUTOMATIC LASER FILTER FOR THE INDIRECT OPHTHALMOSCOPE [J].
GRISOLANO, J ;
PEYMAN, GA .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 1987, 7 (01) :32-33
[4]   CHANGES IN COLOR CONTRAST SENSITIVITY ASSOCIATED WITH OPERATING ARGON-LASERS [J].
GUNDUZ, K ;
ARDEN, GB .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1989, 73 (04) :241-246
[5]   RETINAL SENSITIVITY TO DAMAGE FROM SHORT WAVELENGTH LIGHT [J].
HAM, WT ;
MUELLER, HA ;
SLINEY, DH .
NATURE, 1976, 260 (5547) :153-155
[6]  
HAMPTON GR, 1987, ARCH OPHTHALMOL-CHIC, V105, P132
[7]  
LESPERANCE FA, 1989, OPHTHALMIC LASERS PH, V2, P751
[8]  
MAINSTER MA, 1983, OPHTHALMOLOGY, V90, P973
[9]  
MAINSTER MA, 1989, RETINA, V2, P749
[10]  
MANISTER MA, 1983, OPHTHALMOLOGY, V90, P927