COMPARISON OF PEDIATRIC END-TIDAL CO2 MEASURED WITH NASAL ORAL CANNULA CIRCUIT AND CAPILLARY PCO2

被引:15
作者
ABRAMO, TJ
COWAN, MR
SCOTT, SM
PRIMM, PA
WIEBE, RA
SIGNS, M
机构
[1] Division of Pediatric Emergency Medicine, Department of Pediatrics, University of Texas Southwestern Medical Center at Dallas, Dallas, TX
[2] Division of Pediatric Emergency Medicine, Department of Pediatrics, Eastern Virginia Medical College, Norfolk, VA
关键词
END-TIDAL CARBON DIOXIDE; CAPNOMETRY; ORAL NASAL CANNULA; CAPILLARY PCO2;
D O I
10.1016/0735-6757(95)90236-8
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
This study was designed to determine whether end-tidal carbon dioxide (ETCO2) values obtained by noninvasive oral/nasal cannula circuit with side-stream capnometry correlate reliably with capillary Pco2 (CapCO2) in a pediatric population without cardiopulmonary problems. Each patient was monitored until a reliable 5-minute ETCO2 waveform was obtained. A capillary blood gas sample was drawn while, simultaneously, ETCO2 was recorded. The difference between CapCO2 and ETCO2 levels was tested with a paired t-test at P < .001. The limits of agreement were established with a 95% confidence level. The stability of the measured difference across the range of mean scores (CapCO2 + ETCO2/2), age, and respiratory rate was tested using simple linear regression. Fifty-eight children (23 girls and 35 boys) had mean ETCO2 readings of 33.96 mm Hg (SD 4.26), and mean CapCO2 readings of 35.93 (SD 4.04). A relative average bias of 1.96 with ETCO2 lower than CapCO2 was established with 95% limits of agreement of +/- 5.2 mm Hg (t = 5.71). Variability of difference scores was not related to range of mean scores (r = .08), age (r = .09), or respiratory rate (r = .25). End-tidal CO2 measured by an oral/nasal cannula capnometry circuit is a noninvasive method of assessing indirect measurements of Pco2 in a normal pediatric population.
引用
收藏
页码:30 / 33
页数:4
相关论文
共 17 条
[1]  
Smalhout, A quick guide to capnography and its use in differential diagnosis, (1983)
[2]  
Smalhout, Kalenda, An atlas of capnography, 1, (1975)
[3]  
Bushman, Mangano, Nunn, Monitoring during anesthesia, General Anesthesia, pp. 463-492, (1989)
[4]  
Weil, Bisera, Trevino, Et al., Cardiac output and end-tidal carbon dioxide, Crit Care Med, 13, pp. 907-909, (1985)
[5]  
Gravenstein, Paulus, Hayes, Capnography in clinical practice, (1989)
[6]  
Falk, Rackrow, Weil, End tidal carbon dioxide concentration during cardiopulmonary resuscitation, N Engl J Med, 318, pp. 607-611, (1988)
[7]  
Goldman, A simple, easy, and inexpensive method for monitoring ETCO<sub>2</sub> through nasal cannula, Anesthesiology, 67, (1987)
[8]  
Urmey, Accuracy of expired carbon dioxide partial pressure sampled from a nasal cannula I, Anesthesiology, 68, pp. 959-960, (1988)
[9]  
Dunphy, Accuracy of expired carbon dioxide partial pressure sampled from a nasal cannula II, Anesthesiology, 68, pp. 960-961, (1988)
[10]  
Altman, Bland, Measurements in Medicine The analysis of method comparison studies, The Statistician, 32, pp. 307-317, (1983)