A COMPARISON OF THE COULTER STKS, COULTER S+IV, AND MANUAL ANALYSIS OF WHITE BLOOD-CELL DIFFERENTIAL COUNTS IN A HUMAN IMMUNODEFICIENCY VIRUS-INFECTED POPULATION

被引:11
作者
COHEN, AJ [1 ]
PEERSCHKE, EIB [1 ]
STEIGBIGEL, RT [1 ]
机构
[1] SUNY STONY BROOK,SCH MED,DEPT PATHOL,STONY BROOK,NY 11794
关键词
ACQUIRED IMMUNODEFICIENCY SYNDROME; BLOOD CELL COUNT; CLINICAL TRIALS; ERYTHROCYTE INDEXES; GRANULOCYTES; HUMAN IMMUNODEFICIENCY VIRUS; LYMPHOCYTES;
D O I
10.1093/ajcp/100.6.611
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The white blood cell differential counts obtained by manual analysis and those obtained by automated analysis using Coulter STKS and Coulter S+IV cell counters (Coulter Electronics, Hialeah, FL) in a human immunodeficiency virus (HIV)-infected population were compared. Linear correlation analyses showed the STKS to be less accurate than the S+IV in determinations of the lymphocyte and granulocyte count. This finding was confirmed by difference analyses and by determining the number of measurements that fell outside of 95% confidence limits. Similar analyses were also done on patients not known to be HIV infected, and there was no disparity found between the STKS and S+IV in that group. In the HIV-infected population, the discrepancy between the manual and automated granulocyte counts performed on the STKS, but not on the S+IV, increased with increasing red blood cell mean corpuscular volume. Physicians caring for, and investigators performing studies on, HIV-infected patients should be aware of the potential for falsely low granulocyte numbers when using the differential from the STKS.
引用
收藏
页码:611 / 617
页数:7
相关论文
共 12 条
  • [1] HEMATOLOGIC ABNORMALITIES IN AIDS
    ABOULAFIA, DM
    MITSUYASU, RT
    [J]. HEMATOLOGY-ONCOLOGY CLINICS OF NORTH AMERICA, 1991, 5 (02) : 195 - 214
  • [2] DETECTION OF IMPORTANT ABNORMALITIES OF THE DIFFERENTIAL COUNT USING THE COULTER STKR BLOOD COUNTER
    BARNARD, DF
    BARNARD, SA
    CARTER, AB
    MACHIN, SJ
    PATTERSON, KG
    YARDUMIAN, A
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1989, 42 (07) : 772 - 776
  • [3] AN EVALUATION OF THE COULTER VCS DIFFERENTIAL COUNTER
    BARNARD, DF
    BARNARD, SA
    CARTER, AB
    PATTERSON, KG
    YARDUMIAN, A
    MACHIN, SJ
    [J]. CLINICAL AND LABORATORY HAEMATOLOGY, 1989, 11 (03): : 255 - 266
  • [4] CASTELLA A, 1985, American Journal of Clinical Pathology, V84, P425
  • [5] EVALUATION OF THE COULTER S-PLUS IV 3-PART DIFFERENTIAL IN AN ACUTE CARE HOSPITAL
    GRISWOLD, DJ
    CHAMPAGNE, VD
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1985, 84 (01) : 49 - 57
  • [6] USE OF THE AUTOMATED DIFFERENTIAL LEUKOCYTE COUNT IN PEDIATRIC ONCOLOGY
    KELLEHER, JF
    JANIK, DL
    [J]. JOURNAL OF PEDIATRICS, 1991, 118 (04) : 578 - 580
  • [7] KOEPKE JA, 1985, BLOOD CELLS, V11, P173
  • [8] PIERRE RV, 1991, PRACTICAL LABORATORY, P131
  • [9] ROBERTSON EP, 1992, CLIN LAB HAEMATOL, V14, P53
  • [10] Rumke CL, 1978, DIFFERENTIAL LEUKOCY, P39