COMPARISON OF MEASUREMENT VARIABILITY IN SUBJECTS WITH MODERATE PERIODONTITIS USING A CONVENTIONAL AND CONSTANT FORCE PERIODONTAL PROBE

被引:75
作者
OSBORN, JB [1 ]
STOLTENBERG, JL [1 ]
HUSO, BA [1 ]
AEPPLI, DM [1 ]
PIHLSTROM, BL [1 ]
机构
[1] FLORIDA PROBE CORP,GAINESVILLE,FL
关键词
PERIODONTAL DISEASES; DENTAL INSTRUMENTS;
D O I
10.1902/jop.1992.63.4.283
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
THIS STUDY COMPARED THE intra- and inter-examiner measurement error of the Florida Probe, Florida Disk Probe, and conventional periodontal probe in persons having moderate to severe periodontitis. Single pass measurements using the conventional probe were compared to both single and double pass (corrected) measurements using the Florida Probe and Florida Disk Probe. The mean intra-examiner standard deviations of differences in repeated site probing depth measurements ranged from 0.60 to 0.93 mm for the Florida Probe and from 0.52 to 0.89 mm for the conventional probe. Based on overall averages, the conventional probe measured deeper probing depths than the Florida Probe at the same sites. This difference was especially evident in pockets greater-than-or-equal-to 6 mm in depth. Based on sites, the mean intra-examiner standard deviations of differences in repeated relative attachment level measurements using the Florida Disk Probe ranged from 0.55 to 0.82 mm. The mean standard deviations of site measurements for attachment level obtained with the conventional probe ranged from 0.62 to 1.14 mm. Inter-examiner measurement error was greater than intra-examiner measurement error for each probing method. It was concluded that use of the Florida Probe and Florida Disk Probe may offer significant advantages in reducing measurement error for some clinical examiners. For other examiners, use of the Florida Probes do not necessarily result in less measurement error than use of conventional probing methods.
引用
收藏
页码:283 / 289
页数:7
相关论文
共 18 条
[1]   EFFECT OF TRAINING AND PROBING FORCE ON THE REPRODUCIBILITY OF POCKET DEPTH MEASUREMENTS [J].
ABBAS, F ;
HART, AAM ;
OOSTING, J ;
VANDERVELDEN, U .
JOURNAL OF PERIODONTAL RESEARCH, 1982, 17 (02) :226-234
[2]   REPRODUCIBILITY OF PROBING ATTACHMENT LEVEL MEASUREMENTS [J].
BADERSTEN, A ;
NILVEUS, R ;
EGELBERG, J .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1984, 11 (07) :475-485
[3]  
Fleiss JL., 1986, DESIGN ANAL CLIN EXP, P11
[4]   EVALUATION OF PERIODONTAL PROBING FORCES [J].
FREED, HK ;
GAPPER, RL ;
KALKWARF, KL .
JOURNAL OF PERIODONTOLOGY, 1983, 54 (08) :488-492
[5]   DESCRIPTION AND CLINICAL-EVALUATION OF A NEW COMPUTERIZED PERIODONTAL PROBE - THE FLORIDA PROBE [J].
GIBBS, CH ;
HIRSCHFELD, JW ;
LEE, JG ;
LOW, SB ;
MAGNUSSON, I ;
THOUSAND, RR ;
YERNENI, P ;
CLARK, WB .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1988, 15 (02) :137-144
[6]  
GLAVIND LEIF, 1967, J PERIODONTAL RES, V2, P180, DOI 10.1111/j.1600-0765.1967.tb01887.x
[7]   COMPARISON OF DIFFERENT DATA ANALYSES FOR DETECTING CHANGES IN ATTACHMENT LEVEL [J].
HAFFAJEE, AD ;
SOCRANSKY, SS ;
GOODSON, JM .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1983, 10 (03) :298-310
[8]   COMPARISON OF MANUAL AND PRESSURE-CONTROLLED PERIODONTAL PROBING [J].
KALKWARF, KL ;
KALDAHL, WB ;
PATIL, KD .
JOURNAL OF PERIODONTOLOGY, 1986, 57 (08) :467-471
[9]   ERRORS IN MEASURING PARAMETERS ASSOCIATED WITH PERIODONTAL HEALTH AND DISEASE [J].
KINGMAN, A ;
LOE, H ;
ANERUD, A ;
BOYSEN, H .
JOURNAL OF PERIODONTOLOGY, 1991, 62 (08) :477-486
[10]   CORRELATION BETWEEN ELECTRONIC AND VISUAL READINGS OF POCKET DEPTHS WITH A NEWLY DEVELOPED CONSTANT FORCE PROBE [J].
MAGNUSSON, I ;
FULLER, WW ;
HEINS, PJ ;
RAU, CF ;
GIBBS, CH ;
MARKS, RG ;
CLARK, WB .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1988, 15 (03) :180-184