A REAL-TIME ULTRASOUND TIME-DOMAIN CORRELATION BLOOD FLOWMETER .2. PERFORMANCE AND EXPERIMENTAL-VERIFICATION

被引:16
作者
HEIN, IA [1 ]
OBRIEN, WD [1 ]
机构
[1] UNIV ILLINOIS,DEPT ELECT & COMP ENGN,BIOACOUST RES LAB,URBANA,IL 61801
基金
美国国家卫生研究院;
关键词
D O I
10.1109/58.248223
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The theory and design of a real-time ultrasound time-domain correlation (UTDC) flowmeter employing a high-speed residue number system correlator has been presented in Part I [4] of this work. Part II evaluates the performance of the flowmeter under various How conditions for different correlation parameter settings. Previous work has incorporated a weighted-averaging scheme to estimate the flow velocity over a wide number of echo signal spacings. This research indicates that a linear averaging of 50 adjacent echo signal pairs produces equivalent accuracy and precision at real-time processing speed as compared to previous non-real time systems incorporating weighted averaging.
引用
收藏
页码:776 / 785
页数:10
相关论文
共 13 条
[1]  
Embree P, 1986, THESIS U ILLINOIS UR
[2]   VOLUMETRIC BLOOD-FLOW VIA TIME-DOMAIN CORRELATION - EXPERIMENTAL-VERIFICATION [J].
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :176-189
[3]   FLOW VELOCITY PROFILE VIA TIME-DOMAIN CORRELATION ERROR ANALYSIS AND COMPUTER-SIMULATION [J].
FOSTER, SG ;
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :164-175
[4]  
HEIN IA, 1989, J CARDIOVASC TECHNOL, V8, P339
[5]  
HEIN IA, 1992, IEEE T BIOMED ENG, V39
[6]  
HEIN IA, 1993, IEEE T ULTRASON FERR, V40
[7]  
HEIN IA, 1993, IEEE T ULTRASON FERR, V40, P1
[8]   HEMODYNAMICS OF THE NORMAL HUMAN CAROTID BIFURCATION - INVITRO AND INVIVO STUDIES [J].
KU, DN ;
GIDDENS, DP ;
PHILLIPS, DJ ;
STRANDNESS, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1985, 11 (01) :13-26
[9]  
Nichols WV, 1990, MCDONALDS BLOOD FLOW
[10]  
SMITH NB, 1990 P ULTR S, P1385