3-DIMENSIONAL ECHOCARDIOGRAPHY - IN-VITRO AND IN-VIVO VALIDATION OF LEFT-VENTRICULAR MASS AND COMPARISON WITH CONVENTIONAL ECHOCARDIOGRAPHIC METHODS

被引:102
作者
GOPAL, AS
KELLER, AM
SHEN, ZQ
SAPIN, PM
SCHROEDER, KM
KING, DL
KING, DL
机构
[1] DANBURY HOSP,DIV CARDIOL,DANBURY,CT
[2] UNIV KENTUCKY,DIV CARDIOL,LEXINGTON,KY
[3] FREE UNIV BERLIN,DIV CARDIOL,W-1000 BERLIN,GERMANY
关键词
D O I
10.1016/0735-1097(94)90310-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. This study aimed to validate a method for mass computation in vitro and in vivo and to compare it with conventional methods. Background. Conventional echocardiographic methods of determining left ventricular mass are limited by assumptions of ventricular geometry and image plane positioning. To improve accuracy, we developed a three-dimensional echocardiographic method that uses nonparallel, nonintersecting short-axis planes and a polyhedral surface reconstruction algorithm for mass computation. Methods. Eleven fixed hearts were imaged by three-dimensional echocardiography, and mass was determined in vitro by multiplying the myocardial volume by the density of each heart and comparing it with the true mass. Mass at diastole and systole by three-dimensional echocardiography and magnetic resonance im aging (MRI) was compared in vivo in 15 normal subjects. Ten subjects also underwent imaging by one- and two-dimensional echocardiography, and mass was determined by Penn convention, area-length and truncated ellipsoid algorithms. Results. In vitro results were r = 0.995, SEE 2.91 g, accuracy 3.47%. In vivo interobserver variability for systole and diastole was 16.7% to 27%, 14% to 18.1% and 6.3% to 12.8%, respectively, for one-, two- and three-dimensional echocardiography and was 7.5% for MRI at end-diastole. The latter two agreed closely with regard to diastolic mass (r = 0.895, SEE 11.1 g) and systolic mass (r = 0.926, SEE 9.2 g). These results were significantly better than correlations between MRI and the Penn convention (r = 0.725, SEE 25.6 g for diastole; r = 0.788, SEE 28.7 g for systole), area-length (r = 0.694, SEE 24.2 g for diastole; r = 0.717, SEE 28.2 g for systole) and truncated ellipsoid algorithms (r = 0.687, SEE 21.8 g for diastole; r = 0.710, SEE 24.5 g for systole). Conclusions. Image plane positioning guidance and elimination of geometric assumptions by three-dimensional echocardiography achieve high accuracy for left ventricular mass determination in vitro. It is associated with higher correlations and lower standard errors than conventional methods in vivo.
引用
收藏
页码:504 / 513
页数:10
相关论文
共 37 条
[1]   LEFT-VENTRICULAR MASS AND VOLUME MASS RATIO DETERMINED BY TWO-DIMENSIONAL ECHOCARDIOGRAPHY IN NORMAL ADULTS [J].
BYRD, BF ;
WAHR, D ;
WANG, YS ;
BOUCHARD, A ;
SCHILLER, NB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 6 (05) :1021-1025
[2]  
COLLINS HW, 1989, J AM COLL CARDIOL, V14, P672
[3]   AN ALGORITHM FOR VOLUME ESTIMATION BASED ON POLYHEDRAL APPROXIMATION [J].
COOK, LT ;
COOK, PN ;
LEE, KR ;
BATNITZKY, S ;
WONG, BYS ;
FRITZ, SL ;
OPHIR, J ;
DWYER, SJ ;
BIGONGIARI, LR ;
TEMPLETON, AW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (09) :493-500
[4]   SYSTEM FOR ULTRASONICALLY IMAGING HUMAN HEART IN 3 DIMENSIONS [J].
DEKKER, DL ;
PIZIALI, RL ;
DONG, E .
COMPUTERS AND BIOMEDICAL RESEARCH, 1974, 7 (06) :544-553
[5]   ECHOCARDIOGRAPHIC DETERMINATION OF LEFT-VENTRICULAR MASS IN MAN - ANATOMIC VALIDATION OF METHOD [J].
DEVEREUX, RB ;
REICHEK, N .
CIRCULATION, 1977, 55 (04) :613-618
[6]   ECHOCARDIOGRAPHIC ASSESSMENT OF LEFT-VENTRICULAR HYPERTROPHY - COMPARISON TO NECROPSY FINDINGS [J].
DEVEREUX, RB ;
ALONSO, DR ;
LUTAS, EM ;
GOTTLIEB, GJ ;
CAMPO, E ;
SACHS, I ;
REICHEK, N .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) :450-458
[7]   ACCURATE VOLUME DETERMINATION IN THE ISOLATED EJECTING CANINE LEFT-VENTRICLE BY 2-DIMENSIONAL ECHOCARDIOGRAPHY [J].
EATON, LW ;
MAUGHAN, WL ;
SHOUKAS, AA ;
WEISS, JL .
CIRCULATION, 1979, 60 (02) :320-326
[8]   DYNAMIC 3-DIMENSIONAL ECHOCARDIOGRAPHIC RECONSTRUCTION OF THE INTACT HUMAN LEFT-VENTRICLE - TECHNIQUE AND INITIAL OBSERVATIONS IN PATIENTS [J].
GEISER, EA ;
ARIET, M ;
CONETTA, DA ;
LUPKIEWICZ, SM ;
CHRISTIE, LG ;
CONTI, CR .
AMERICAN HEART JOURNAL, 1982, 103 (06) :1056-1065
[9]   3-DIMENSIONAL RECONSTRUCTION OF ECHOCARDIOGRAPHIC IMAGES USING THE ROTATION METHOD [J].
GHOSH, A ;
NANDA, NC ;
MAURER, G .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (06) :655-661
[10]  
Gopal A S, 1992, J Am Soc Echocardiogr, V5, P115