CORRELATION BETWEEN THE RESISTANCE RATIOS OF PLATINUM RESISTANCE THERMOMETERS AT THE MELTING-POINT OF GALLIUM AND THE TRIPLE POINT OF MERCURY

被引:16
作者
SINGH, YP
MAAS, H
EDLER, F
ZAIDI, ZH
机构
[1] PHYS TECH BUNDESANSTALT,D-10587 BERLIN,GERMANY
[2] JAMIA MILLIA ISLAMIA UNIV,NEW DELHI 110025,INDIA
关键词
D O I
10.1088/0026-1394/31/1/010
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A set of resistance ratios (W) for platinum resistance thermometers was obtained at the triple point of Hg and the melting point of Ga in order to study their relationship. It was found that using measured values for one of the fixed points, a linear equation will predict the value of the other. These measurements also indicate that the fixed points of Hg and of Ga are inconsistent by about 1,5 mK in the sense that either the melting point of Ga or the triple point of Hg was assigned too high a value on the ITS-90.
引用
收藏
页码:49 / 50
页数:2