IN-VITRO TORQUE-DEFORMATION CHARACTERISTICS OF ORTHODONTIC POLYCARBONATE BRACKETS

被引:46
作者
FELDNER, JC
SARKAR, NK
SHERIDAN, JJ
LANCASTER, DM
机构
[1] LOUISIANA STATE UNIV,SCH DENT,DEPT BIOMAT,NEW ORLEANS,LA 70119
[2] LOUISIANA STATE UNIV,SCH DENT,DEPT ORTHODONT,NEW ORLEANS,LA 70119
[3] LOUISIANA STATE UNIV,SCH DENT,OFF INSTRUCT SERV,NEW ORLEANS,LA 70119
关键词
D O I
10.1016/S0889-5406(94)70046-X
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to investigate the torque-deformation characteristics of the following four types of polycarbonate brackets: (1) pure polycarbonate, PPC (anterior Miura, RMO, Denver, Cole.); (2) ceramic reinforced polycarbonate, CRPC (Silkon bracket, American, Sheboygan, Wis.), (3) metal slot reinforced polycarbonate, MRPC (Plastic bracket, Tella Tech, Miami, Fla.), and (4) metal slot and ceramic reinforced polycarbonate, MCRPC (Spirit, Ormco, Glendora, Calif.). A stainless steel bracket, (Mini Diamond, Ormco, Glendora, Calif.), was used as a control. Ten brackets of each type were tested. Each bracket was bonded to a porcelain tooth and engaged in a torquemeter. The tooth-bracket assembly was made stationary by embeding it in die stone. Torsion was applied to the bracket at 4 degrees per minute and the resultant torque (grams . centimeters) and deformation (degree) were measured. For optimum labiolingual tooth movement for a maxillary incisor at 175 grams . centimeters, the amount of angular deflection necessary for the different polycarbonate brackets was the following: (a) 15 degrees for MRPC, (b) 17 degrees for MCRPC, (c) 24 degrees for CRPC, and (d) >30 degrees for PPC. The amount of deformation at this deflection was the least for MRCP followed by MCRPC, CRCP, and PPC. When compared with the stainless steel bracket, all polycarbonate brackets showed significantly (p < 0.0001) higher deformation and lower torque. Within the polycarbonate group, there was a significant difference (p < 0.0001) between each bracket for both measurements. The MRPC produced the highest torque and lowest deformation values followed by the MCRPC, CRCP, and PPC. It appears that only the metal slot reinforced brackets are clinically capable of torquing teeth sufficiently.
引用
收藏
页码:265 / 272
页数:8
相关论文
共 23 条
[1]  
Aird J C, 1987, Br J Orthod, V14, P191
[2]   EVALUATION OF FRICTION BETWEEN CERAMIC BRACKETS AND ORTHODONTIC WIRES OF 4 ALLOYS [J].
ANGOLKAR, PV ;
KAPILA, S ;
DUNCANSON, MG ;
NANDA, RS .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1990, 98 (06) :499-506
[3]  
BANTLEON H P, 1988, Fortschritte der Kieferorthopaedie, V49, P203, DOI 10.1007/BF02163378
[4]  
BURSTONE CJ, 1966, ANGLE ORTHOD, V36, P99
[5]  
CHOI LK, 1989, THESIS LOMA LINDA U
[6]   BOND STRENGTH OF ORTHODONTIC DIRECT-BONDING CEMENT PLASTIC BRACKET SYSTEMS INVITRO [J].
DEPULIDO, LG ;
POWERS, JM .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1983, 83 (02) :124-130
[7]   LOAD-DEFORMATION CHARACTERISTICS OF POLYCARBONATE ORTHODONTIC BRACKETS [J].
DOBRIN, RJ ;
KAMEL, IL ;
MUSICH, DR .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1975, 67 (01) :24-33
[8]   ENAMEL WEAR CAUSED BY CERAMIC BRACKETS [J].
DOUGLASS, JB .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1989, 95 (02) :96-98
[9]  
Gottlieb E L, 1991, J Clin Orthod, V25, P145
[10]   FRACTURE-RESISTANCE OF CERAMIC BRACKETS DURING ARCH WIRE TORSION [J].
HOLT, MH ;
NANDA, RS ;
DUNCANSON, MG .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1991, 99 (04) :287-293