SENSITIVE RF-SQUIDS AND MAGNETOMETERS OPERATING AT 77 K

被引:72
作者
Zhang, Y. [1 ]
Mueck, M. [1 ]
Herrmann, K. [1 ]
Schubert, J. [1 ]
Zander, W. [1 ]
Braginski, A. I. [1 ]
Heiden, C. [1 ]
机构
[1] Forschungszentrum Julich, Inst Schicht & Ionentech, POB 1913,W 5170, D-52425 Julich, Germany
关键词
10;
D O I
10.1109/77.233955
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Large 6 X 6 mm(2) and 8 X 8 mm(2) flux-focussing washer structures with inductances L-S between 25 pH and 500 pH were fabricated from epitaxial, c- axis YBa2Cu3O7 films. Double step- edge junctions having a low l/ f noise were incorporated in these washers to form rf- SQUIDS. A high tank circuit frequency near 150 MHz was chosen to reduce noise and improve sensitivity of the SQUIDS. At 77K, a transfer function exceeding 100 mu V Phi(0) and white flux noise of 3 X 10(-5) Phi(0)/root Hz down to below 0.3 Hz, were attained with L-S = 25 pH. A magnetometer with L-S = 190 pH was demonstrated. The best magnetic field and energy resolutions were 170 fT/root Hz and 5X10(-29) J/ Hz respectively over a signal frequency range to below 1 Hz. The magnetometer was used to record human magnetocardiograms, and, for the first time, auditory evoked human brain responses.
引用
收藏
页码:2465 / 2468
页数:4
相关论文
共 9 条
[1]  
Cui G., 1991, SUPERCOND SCI TECH, V4, P130
[2]  
Elbert T., 1992, J ELECTROENCEP UNPUB
[3]   CHARACTERIZATION OF YBA2CU3O7 STEP-EDGE JOSEPHSON-JUNCTIONS [J].
HERRMAN, K ;
ZHANG, Y ;
MUCK, HM ;
SCHUBERT, J ;
ZANDER, W ;
BRAGINSKI, AI .
SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1991, 4 (11) :583-586
[4]   NOISE IN RF SQUID [J].
JACKEL, LD ;
BUHRMAN, RA .
JOURNAL OF LOW TEMPERATURE PHYSICS, 1975, 19 (3-4) :201-246
[5]  
KETCHEN MB, 1985, SQUID 85, P865
[6]   NOISE IN SUPERCONDUCTING QUANTUM FLUX DETECTOR [J].
KURKIJARVI, J .
JOURNAL OF APPLIED PHYSICS, 1973, 44 (08) :3729-3733
[7]   THIN-FILM YBCO MAGNETOMETER [J].
MIKLICH, AH ;
WELLSTOOD, FC ;
KINGSTON, JJ ;
CLARKE, J ;
COLCLOUGH, MS ;
CHAR, K ;
ZAHARCHUK, G .
NATURE, 1991, 352 (6335) :482-483
[8]   LOW-NOISE YBA2CU3O7 RF SQUID MAGNETOMETER [J].
ZHANG, Y ;
MUCK, HM ;
HERRMANN, K ;
SCHUBERT, J ;
ZANDER, W ;
BRAGINSKI, AI ;
HEIDEN, C .
APPLIED PHYSICS LETTERS, 1992, 60 (05) :645-647
[9]  
Zhang Y., P ISBET 199 IN PRESS