HEMODYNAMIC EVALUATION OF THE CARBOMEDICS PROSTHETIC HEART-VALVE IN THE AORTIC POSITION - COMPARISON OF NONINVASIVE AND INVASIVE TECHNIQUES

被引:58
作者
IHLEN, H [1 ]
MOLSTAD, P [1 ]
SIMONSEN, S [1 ]
VATNE, K [1 ]
OVRUM, E [1 ]
GEIRAN, O [1 ]
LAAKE, P [1 ]
FROYSAKER, T [1 ]
机构
[1] UNIV HOSP OSLO,RIKSHOSP,SURG DEPT A,OSLO,NORWAY
关键词
D O I
10.1016/0002-8703(92)90759-O
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Seventy-three patients with a CarboMedics aortic bileaflet valve prosthesis were examined by Doppler ultrasonography, and 27 of them were also assessed by transseptal catheterization. The ultrasonic mean systolic gradient was 17.1 +/- 5.6 mm Hg for valve size 19 mm, falling gradually with increasing valve size to 6.8 +/- 2.5 mm Hg for size 27 mm. The catheter mean systolic gradient was consistently smaller than the ultrasonic gradient (4.3 +/- 4.8 mm Hg), but Tobit regression analysis showed a significant association between the two methods. In all patients both methods revealed negligible to small amounts of retrograde leakage, which is assumed to be a normal finding for this valve. The effective flow areas of the valves calculated from the ultrasonic data were similar to the in vitro calculated flow areas. The hemodynamic potential of this valve is therefore completely utilized in vivo. The effective orifice area corrected for body surface area increased with increasing valve size, which demonstrates a moderate valve-patient mismatch.
引用
收藏
页码:151 / 159
页数:9
相关论文
共 23 条
[1]   COLOR FLOW DOPPLER EVALUATION OF ST-JUDE MEDICAL PROSTHETIC VALVES [J].
ALAM, M ;
ROSMAN, HS ;
MCBROOM, D ;
GRAHAM, L ;
MAGILLIGAN, DJ ;
KHAJA, F ;
STEIN, PD .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (19) :1387-1389
[2]   DISCREPANCIES BETWEEN DOPPLER AND CATHETER GRADIENTS IN AORTIC PROSTHETIC VALVES INVITRO - A MANIFESTATION OF LOCALIZED GRADIENTS AND PRESSURE RECOVERY [J].
BAUMGARTNER, H ;
KHAN, S ;
DEROBERTIS, M ;
CZER, L ;
MAURER, G .
CIRCULATION, 1990, 82 (04) :1467-1475
[3]   CONTINUOUS WAVE DOPPLER ECHOCARDIOGRAPHIC MEASUREMENT OF PROSTHETIC VALVE GRADIENTS - A SIMULTANEOUS DOPPLER-CATHETER CORRELATIVE STUDY [J].
BURSTOW, DJ ;
NISHIMURA, RA ;
BAILEY, KR ;
REEDER, GS ;
HOLMES, DR ;
SEWARD, JB ;
TAJIK, AJ .
CIRCULATION, 1989, 80 (03) :504-514
[4]   INADEQUACY OF THE GORLIN FORMULA FOR PREDICTING PROSTHETIC VALVE AREA [J].
CANNON, SR ;
RICHARDS, KL ;
CRAWFORD, MH ;
FOLLAND, ED ;
PIERPONT, G ;
SETHI, GK ;
HAMMERMEISTER, KE .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (01) :113-116
[5]   A 6 YEAR EXPERIENCE WITH THE ST-JUDE MEDICAL VALVE - HEMODYNAMIC PERFORMANCE, SURGICAL RESULTS, BIOCOMPATIBILITY AND FOLLOW-UP [J].
CZER, LSC ;
MATLOFF, J ;
CHAUX, A ;
DEROBERTIS, M ;
YOGANATHAN, A ;
GRAY, RJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 6 (04) :904-912
[6]  
DENNIG K, 1986, HERZ, V11, P341
[7]   VALIDATION AND APPLICATIONS OF INDEXED AORTIC PROSTHETIC VALVE AREAS CALCULATED BY DOPPLER ECHOCARDIOGRAPHY [J].
DUMESNIL, JG ;
HONOS, GN ;
LEMIEUX, M ;
BEAUCHEMIN, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :637-643
[8]  
EIE H, 1970, NORWEGIAN MONOGRAPHS, P46
[9]   HYDRAULIC FORMULA FOR CALCULATION OF THE AREA OF THE STENOTIC MITRAL VALVE, OTHER CARDIAC VALVES, AND CENTRAL CIRCULATORY SHUNTS .1. [J].
GORLIN, R ;
GORLIN, SG .
AMERICAN HEART JOURNAL, 1951, 41 (01) :1-29
[10]  
HATLE L, 1985, DOPPLER ULTRASOUND C, P22