EDUCATION AND ENGINEERING SOLUTIONS FOR POTENTIAL PROBLEMS WITH LAPAROSCOPIC MONOPOLAR ELECTROSURGERY

被引:76
作者
VOYLES, CR
TUCKER, RD
机构
[1] UNIV MISSISSIPPI,DEPT SURG,JACKSON,MS 39216
[2] UNIV IOWA,DEPT PATHOL,IOWA CITY,IA 52242
关键词
D O I
10.1016/S0002-9610(05)80648-8
中图分类号
R61 [外科手术学];
学科分类号
摘要
The potential problems of monopolar electrosurgery relate to unrecognized energy transfer ("stray current") outside the view of the laparoscope. Mechanisms of stray current and unrecognized tissue injury include: (1) insulation breaks in electrodes; (2) capacitive coupling, or induced currents through the intact insulation of the active electrode to surrounding cannulas or other instruments; and (3) direct coupling (or unintended contact) between the active electrode and other metal instruments or cannulas within the abdomen. Capacitive coupling poses the greatest risk for injury when the outer conductor (trocar cannula or irrigation cannula) is electrically isolated from the abdominal wall by a plastic nonconductor. Capacitive coupling is increased by the coagulation mode (versus, cut), open circuit (versus tissue contact with the electrode), 5-mm cannulas (versus 11 mm), and higher voltage generators. The safety of electrosurgery can be enhanced by surgical education regarding the biophysics of radio frequency electrical energy, technical choices in instruments using all-metal cannula systems, and engineering developments with a dynamically monitored system for insulation failure and capacitive coupling.
引用
收藏
页码:57 / 62
页数:6
相关论文
共 19 条
[1]  
LEVY BS, 1985, J REPROD MED, V30, P168
[2]  
ORGAN CH, 1991, JAMA-J AM MED ASSOC, V265, P3172
[3]   DEATHS ASSOCIATED WITH LAPAROSCOPIC STERILIZATION BY UNIPOLAR ELECTROCOAGULATING DEVICES, 1978 AND 1979 [J].
PETERSON, HB ;
ORY, HW ;
GREENSPAN, JR ;
TYLER, CW .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1981, 139 (02) :141-143
[4]  
SCHWIMMER WB, 1974, OBSTET GYNECOL, V44, P526
[5]  
SERWAY RA, 1990, PHYSICS SCI ENG, P714
[6]  
SINGH KB, 1977, NEW YORK STATE J MED, V77, P190
[7]  
SODERSTROM RM, 1974, JAMA-J AM MED ASSOC, V227, P1261, DOI 10.1001/jama.1974.03230240019011
[8]  
SODERSTROM RM, 1980, CONTEMP OBSTET GYNEC, V16, P121
[9]  
SODERSTROM RM, 1990, CONT OB GYN, V35, P35
[10]   POSSIBLE HAZARD OF LAPAROSCOPIC STERILIZATION [J].
STEWART, KS ;
PEARSON, JF ;
DOCKER, MF ;
HARVEY, LP ;
RUSHTON, DI .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1973, 115 (08) :1154-1157