EVALUATION AND COMPARISON OF THE TDXII, STRATUS, AND OPUS DIGOXIN ASSAYS

被引:14
作者
DODDS, HM [1 ]
NORRIS, RLG [1 ]
JOHNSON, AG [1 ]
POND, SM [1 ]
机构
[1] PRINCESS ALEXANDRA HOSP, DEPT CLIN MED, WOOLLOONGABBA, QLD 4102, AUSTRALIA
关键词
SERUM DIGOXIN; DIGOXIN-LIKE IMMUNOREACTIVE SUBSTANCES; DIGOXIN FAB ANTIBODY; AUTOMATED IMMUNOASSAYS;
D O I
10.1097/00007691-199502000-00011
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Three automated immunoassays for digoxin in serum were evaluated-Abbott TDxII, Baxter Stratus, and Behring OPUS. The accuracy and precision of the assays were assessed by weighed-in controls and an external quality control program. Coefficients of variation of all methods in serum were less than or equal to 10% at weighed-in concentrations of digoxin of 1 and 2.5 mu g/L. Accuracy relative to weighed-in concentrations of 1 and 2.5 mu g/L ranged from 98 to 126% for all methods. Comparative results from patient samples showed little difference between the TDxII and Stratus and a greater difference observed between the TDxII and OPUS assays. The detection of digoxin-free samples containing digoxin-like immunoreactive substances (DLIS) in neonatal cord blood, pregnant patients, and liver and renal recipients by each assay was then assessed. The TDxII exhibited the highest incidence of DLIS. This is evident in neonatal cord blood in which 40.4% of samples tested positive. In comparison, the extent of DLIS detected by Stratus was less and OPUS exhibited no DLIS in any of the groups studied. A case study of a patient treated with anti-digoxin Fab fragments (Digibind) also was included for analysis by each method. Fourteen hours after Digibind administration, the TDxII registered a digoxin concentration of 49.5 mu g/L compared with 3.73, 1.80, and 2.49 mu g/L for Stratus, OPUS, and ultrafiltered TDxII methods, respectively. The results indicate that to determine the concentration of digoxin after the administration of Digibind, the OPUS or fluorescence polarization immunoassay (FPIA)-ultrafiltered samples by TDxII are the assays of choice.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 28 条
[1]   TREATMENT OF 150 CASES OF LIFE-THREATENING DIGITALIS INTOXICATION WITH DIGOXIN-SPECIFIC FAB ANTIBODY FRAGMENTS - FINAL REPORT OF A MULTICENTER STUDY [J].
ANTMAN, EM ;
WENGER, TL ;
BUTLER, VP ;
HABER, E ;
SMITH, TW .
CIRCULATION, 1990, 81 (06) :1744-1752
[2]  
BALZAN S, 1984, CLIN CHEM, V30, P450
[3]  
BEYERS AD, 1984, S AFR MED J, V65, P878
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   INACCURACIES IN DIGOXIN MEASUREMENT [J].
COOK, JD ;
KOCH, TR ;
COOK, MS ;
KNOBLOCK, EC .
CLINICAL BIOCHEMISTRY, 1988, 21 (06) :353-357
[6]   DIGOXIN IMMUNOREACTIVITY IN CORD AND MATERNAL SERUM AND PLACENTAL EXTRACTS - PARTIAL CHARACTERIZATION OF IMMUNOREACTIVE SUBSTANCES BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND INHIBITION OF NA+, K+-ATPASE [J].
DIAMANDIS, EP ;
PAPANASTASIOUDIAMANDI, A ;
SOLDIN, SJ .
CLINICAL BIOCHEMISTRY, 1985, 18 (01) :48-55
[7]  
GAULT MH, 1986, CLIN CHEM, V32, P2000
[8]  
GRAVES SW, 1986, CLIN CHEM, V32, P1506
[9]   AN ENDOGENOUS DIGOXIN-LIKE SUBSTANCE IN PATIENTS WITH RENAL IMPAIRMENT [J].
GRAVES, SW ;
BROWN, B ;
VALDES, R .
ANNALS OF INTERNAL MEDICINE, 1983, 99 (05) :604-608
[10]  
HURSTING MJ, 1987, CLIN CHEM, V33, P1652