NUMERICAL-SIMULATION OF ANNULAR-PHASED ARRAYS OF DIPOLES FOR HYPERTHERMIA OF DEEP-SEATED TUMORS

被引:34
作者
CHEN, JY
GANDHI, OP
机构
[1] Department of Electrical Engineering, University of Utah, Salt Lake City
关键词
D O I
10.1109/10.125005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have used the finite-difference time domain (FDTD) method to calculate the SAR distributions from an annular-phased array of eight dipole antennas coupled through water "boluses" in anatomically based three-dimensional models of the human body. We evaluated the effect of tapered bolus chambers, frequency (100-120 MHz), dipole length (17-30 cm), and phase and amplitude of power to the various dipoles on the ability to focus energy in the region of deep-seated tumors in the prostate and the liver. Assuming tumor conductivity and permittivity to be similar or slightly higher than surrounding normal tissues, calculations indicate that adjustment of the noted parameters should result in considerable improvement in focusing of SAR distributions in tumor-bearing regions. If such calculations can be shown to correctly predict empirical measurements from complex inhomogeneous (although not necessarily anatomically correct) phantoms, they may be useful for hyperthermia treatment planning based on patient-specific anatomic models.
引用
收藏
页码:209 / 216
页数:8
相关论文
共 24 条
[1]   ELECTROMAGNETIC DEPOSITION IN AN ANATOMICALLY BASED MODEL OF MAN FOR LEAKAGE FIELDS OF A PARALLEL-PLATE DIELECTRIC HEATER [J].
CHEN, JY ;
GANDHI, OP .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1989, 37 (01) :174-180
[2]   RF CURRENTS INDUCED IN AN ANATOMICALLY-BASED MODEL OF A HUMAN FOR PLANE-WAVE EXPOSURES (20-100-MHZ) [J].
CHEN, JY ;
GANDHI, OP .
HEALTH PHYSICS, 1989, 57 (01) :89-98
[3]   HYPERTHERMIC THERAPY OF DEEP SEATED TUMORS - COMPARISON OF THE HEATING EFFICIENCIES OF AN ANNULAR ARRAY APPLICATOR AND A CAPACITIVELY COUPLED RADIOFREQUENCY SYSTEM [J].
EGAWA, S ;
TSUKIYAMA, I ;
AKINE, Y ;
KAJIURA, Y ;
OGINO, T ;
YAMASHITA, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (03) :521-528
[4]  
Eycleshymer AC, 1911, CROSS SECTION ANATOM
[5]  
FOSTER KR, 1981, J MICROWAVE POWER EE, V16, P107
[6]   ABERRANT HEATING - A PROBLEM IN REGIONAL HYPERTHERMIA [J].
HAGMANN, MJ ;
LEVIN, RL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (04) :405-411
[7]   REGIONAL HYPERTHERMIA FOR EXTENSIVE PELVIC TUMORS USING AN ANNULAR PHASED-ARRAY APPLICATOR - A FEASIBILITY STUDY [J].
HOWARD, GCW ;
SATHIASEELAN, V ;
KING, GA ;
DIXON, AK ;
ANDERSON, A ;
BLEEHEN, NM .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (708) :1195-1201
[8]   NUMERICAL-CALCULATIONS OF THE TEMPERATURE DISTRIBUTION IN REALISTIC CROSS-SECTIONS OF THE HUMAN-BODY [J].
ISKANDER, MF ;
KHOSHDELMILANI, O .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (10) :1907-1912
[9]  
Issels R D, 1988, Recent Results Cancer Res, V107, P236
[10]   HYPERTHERMIC ONCOLOGY - CURRENT BIOLOGY, PHYSICS AND CLINICAL-RESULTS [J].
MEYER, JL ;
KAPP, DS ;
FESSENDEN, P ;
HAHN, GH .
PHARMACOLOGY & THERAPEUTICS, 1989, 42 (02) :251-288