The validation of the ultrasound time-domain correlation method of measuring blood flow has required the development of a flexible blood flow phantom capable of generating predictable flow profiles under a wide variety of conditions. The purpose of the phantom is to generate flow with well-known flow properties and not to mimic actual in vivo vessels. This paper describes a flow phantom which can independently generate both constant and pulsatile flow over a wide range of flow rates with a spatially fully developed laminar flow profile. It incorporates a computer-controlled pulsatile pump, which can produce different temporal pulsatile waveforms. The flow phantom also supports multiple vessels, different vessel sizes, as well as different attenuating media. The fluid most commonly used in the phantom is Sephadex mixed with water, and the probability density function of ultrasound reflected from Sephadex is experimentally determined and compared with that of blood. Examples of different constant and pulsatile flow experiments using the phantom are presented.
引用
收藏
页码:1111 / 1122
页数:12
相关论文
共 30 条
[1]  
Alberts B., 1983, MOL BIOL CELL
[2]   THEORETICAL-STUDY OF THE SCATTERING OF ULTRASOUND FROM BLOOD [J].
ANGELSEN, BAJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (02) :61-67
[3]  
[Anonymous], 1989, HEMODYNAMICS
[4]  
BRODY WR, 1974, IEEE T BIO-MED ENG, VBM21, P183, DOI 10.1109/TBME.1974.324381
[5]   AN INVITRO MODEL AND ITS APPLICATION FOR THE STUDY OF CAROTID DOPPLER SPECTRAL BROADENING [J].
DOUVILLE, Y ;
JOHNSTON, KW ;
KASSAM, M ;
ZUECH, P ;
COBBOLD, RSC ;
JARES, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1983, 9 (04) :347-356
[6]   VOLUMETRIC BLOOD-FLOW VIA TIME-DOMAIN CORRELATION - EXPERIMENTAL-VERIFICATION [J].
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :176-189
[7]   FLOW VELOCITY PROFILE VIA TIME-DOMAIN CORRELATION ERROR ANALYSIS AND COMPUTER-SIMULATION [J].
FOSTER, SG ;
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :164-175
[8]   TURBULENCE MEASUREMENTS WITH PULSED DOPPLER ULTRASOUND EMPLOYING A FREQUENCY TRACKING METHOD [J].
GIDDENS, DP ;
KHALIFA, AMA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (04) :427-437
[9]  
HEIN IA, 1989, J CARDIOVASC TECHNOL, V8, P339
[10]  
HEIN IA, 1989, 1989 IEEE ULTR S P M, P881