DELAYED DETECTION OF HYPOXIC EVENTS BY PULSE OXIMETERS - COMPUTER-SIMULATIONS

被引:21
作者
VERHOEFF, F [1 ]
SYKES, MK [1 ]
机构
[1] RADCLIFFE INFIRM,NUFFIELD DEPT ANAESTHET,OXFORD OX2 6HE,ENGLAND
关键词
Measurement techniques; pulse oximetry; Oxygen; saturation;
D O I
10.1111/j.1365-2044.1990.tb14271.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
There is a variable delay between a reduction in alveolar Po2 and the decrease in arterial oxygen saturation recorded on a pulse oximeter. The decrease in arterial oxygen saturation in response to disconnexion of a paralysed patient from the breathing system, oxygen supply failure with continued mechanical ventilation and disconnexion of the fresh gas supply to Mapleson D and circle absorption breathing systems were studied by simulations on the MacPuf computer model of the cardiorespiratory system. The simulations revealed that there were marked differences between the rate of arterial desaturation which resulted from each of the three types of oxygen supply failure and that arterial oxygen saturation may reach dangerous levels before a pulse oximeter alarm is activated. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:103 / 109
页数:7
相关论文
共 6 条
[1]  
Dickinson C., 1977, COMPUTER MODEL HUMAN
[2]  
EICHHORN JH, 1986, JAMA-J AM MED ASSOC, V256, P1017
[3]   ACCURACY OF RESPONSE OF 6 PULSE OXIMETERS TO PROFOUND HYPOXIA [J].
SEVERINGHAUS, JW ;
NAIFEH, KH .
ANESTHESIOLOGY, 1987, 67 (04) :551-558
[4]   ESSENTIAL MONITORING [J].
SYKES, MK .
BRITISH JOURNAL OF ANAESTHESIA, 1987, 59 (07) :901-912
[5]   COMPARISON OF PULSE OXIMETERS - EFFECTS OF VASOCONSTRICTION AND VENOUS ENGORGEMENT [J].
WILKINS, CJ ;
MOORES, M ;
HANNING, CD .
BRITISH JOURNAL OF ANAESTHESIA, 1989, 62 (04) :439-444
[6]  
1988, RECOMMENDATIONS STAN