PRACTICAL LOW-COST STAND SIT SYSTEM FOR MID-THORACIC PARAPLEGICS

被引:15
作者
EWINS, DJ [1 ]
TAYLOR, PN [1 ]
CROOK, SE [1 ]
LIPCZYNSKI, RT [1 ]
SWAIN, ID [1 ]
机构
[1] ODSTOCK HOSP,DEPT MED PHYS,SALISBURY SP2 8BJ,WILTS,ENGLAND
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1988年 / 10卷 / 02期
关键词
BIOMECHANICS - Musculoskeletal Systems - BIOMEDICAL EQUIPMENT - Wheelchairs - COMPUTER SOFTWARE - Medical Applications - PROSTHETICS - Sensory Feedback;
D O I
10.1016/0141-5425(88)90098-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of functional electrical stimulation (FES) to enable paraplegics to stand is not new or indeed difficult to undertake under laboratory conditions. However, there are substantial problems to overcome before such systems can be used routinely by patients without professional supervision. The overriding consideration has to be one of safety, i. e. the system must be 'fail safe'. Second, the system must be quick and easy to use in a wide variety of locations, otherwise it will not provide any increase in function. Finally, it must be inexpensive enough to be available to a large number of paraplegics. The primary aim of our work was to provide such a system to enable mid-thoracic lesion paraplegics to stand wherever they wish. This involved the development of a microprocessor-based stimulator to enable the stimulating envelope to be individually tailored to a given patient's requirements and the provision of closed loop control to minimize fatigue. A folding standing frame was also designed which replaces the arm rests on a standard wheelchair. Using this system, the user is able to stand within 30 //s of stopping and can remain standing for up to 10 min. Cosmetic calipers (knee-ankle-foot orothoses) are also being used for paraplegics who require to stand for longer periods.
引用
收藏
页码:184 / 188
页数:5
相关论文
共 11 条
[1]   THE USE OF A 4-CHANNEL ELECTRICAL STIMULATOR AS AN AMBULATORY AID FOR PARAPLEGIC PATIENTS [J].
BAJD, T ;
KRALJ, A ;
TURK, R ;
BENKO, H ;
SEGA, J .
PHYSICAL THERAPY, 1983, 63 (07) :1116-1120
[2]   USE OF A 2-CHANNEL FUNCTIONAL ELECTRICAL STIMULATOR TO STAND PARAPLEGIC PATIENTS [J].
BAJD, T ;
KRALJ, A ;
SEGA, J ;
TURK, R ;
BENKO, H ;
STROJNIK, P .
PHYSICAL THERAPY, 1981, 61 (04) :526-527
[3]   ADAPTIVE-CONTROL OF ELECTRICALLY STIMULATED MUSCLE [J].
BERNOTAS, LA ;
CRAGO, PE ;
CHIZECK, HJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (02) :140-147
[4]  
BRAY C, 1983, ADV USER GUIDE BBC C
[5]   CLOSED-LOOP CONTROL OF FORCE DURING ELECTRICAL-STIMULATION OF MUSCLE [J].
CRAGO, PE ;
MORTIMER, JT ;
PECKHAM, PH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (06) :306-311
[6]  
CRAGO PE, 1983, ENG MED BIOL MAG SEP, P32
[7]  
CRAGO PE, 1982, SOC NEUR ABSTR, V8, P956
[8]  
JANNSON LA, 1975, 24TP14421 MULL TECHN
[9]  
OPIE C, 1984, INTERFACING BBC MICR
[10]   FEEDBACK-CONTROL SYSTEM FOR WALKING IN MAN [J].
PETROFSKY, JS ;
PHILLIPS, CA ;
HEATON, HH .
COMPUTERS IN BIOLOGY AND MEDICINE, 1984, 14 (02) :135-149