FACTORS AFFECTING SOLID-PHASE EXTRACTION OF SEMIVOLATILE ORGANIC POLLUTANTS FROM ACIDIC INDUSTRIAL EFFLUENT FOR ANALYSIS BY GAS-CHROMATOGRAPHY

被引:17
作者
DEANS, IS
DAVIDSON, CM
LITTLEJOHN, D
BROWN, I
机构
[1] UNIV STRATHCLYDE,DEPT PURE & APPL CHEM,295 CATHEDRAL ST,GLASGOW G1 1XL,SCOTLAND
[2] ZENECA FINE CHEM MFG ORG,GRANGEMOUTH FK3 8XG,STIRLING,SCOTLAND
关键词
SOLID-PHASE EXTRACTION; GAS CHROMATOGRAPHY; WASTE CHEMICAL EFFLUENT; TOLUENE; XYLENE AND CHLOROBENZENES;
D O I
10.1039/an9931801375
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method was developed for the solid-phase extraction of toluene, chlorobenzene, xylene, dichlorobenzene and trichlorobenzene from acidic, aqueous industrial effluent, prior to quantification by gas chromatography (GC). A variety of sorbent-solvent systems were investigated and the highest analyte recoveries were obtained with a C18 sorbent and chloroform. Dichloromethane can also be used as an alternative eluent. Up to 100 ml of effluent were passed through the cartridges, containing 500 mg of C18 sorbent, without degradation of the analyte recoveries. As the analytes were completely eluted with 1 ml of chloroform, preconcentration factors of up to 100-fold were achieved. To obtain close to quantitative (100%) recovery of the analytes from samples it is necessary to prevent evaporative losses which occur (a) during storage and sampling, if a headspace is created above the liquid, (b) when sorbent drying times >2 min are used and (c) if there is a significant delay (e.g., 45 min) between drying and elution of the analytes. Accordingly, each cartridge should be loaded and eluted in series to achieve maximum recoveries. The GC method detection limits (without preconcentration) were 0.18-0.33 mug ml-1, which are adequate for the determination of the analytes at concentrations typical of the effluent studied.
引用
收藏
页码:1375 / 1382
页数:8
相关论文
共 14 条
[1]   SOLID-PHASE MICROEXTRACTION WITH THERMAL-DESORPTION USING FUSED-SILICA OPTICAL FIBERS [J].
ARTHUR, CL ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1990, 62 (19) :2145-2148
[2]   AUTOMATION AND OPTIMIZATION OF SOLID-PHASE MICROEXTRACTION [J].
ARTHUR, CL ;
KILLAM, LM ;
BUCHHOLZ, KD ;
PAWLISZYN, J ;
BERG, JR .
ANALYTICAL CHEMISTRY, 1992, 64 (17) :1960-1966
[3]  
BENJAMIN J, 1985, 2ND P INT S SAMPL PR
[4]   DETERMINATION OF VOLATILE ORGANICS IN DRINKING-WATER WITH USEPA METHOD-524.2 AND THE ION TRAP DETECTOR [J].
EICHELBERGER, JW ;
BELLAR, TA ;
DONNELLY, JP ;
BUDDE, WL .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1990, 28 (09) :460-467
[5]  
HAM RK, 1989, MOD CAST JUL, P27
[6]  
LISKA I, 1989, HRC-J HIGH RES CHROM, V12, P577
[7]  
Loconto P, 1991, LC GC EUR, V4, P10
[8]   DYNAMICS OF ORGANIC-COMPOUND EXTRACTION FROM WATER USING LIQUID-COATED FUSED-SILICA FIBERS [J].
LOUCH, D ;
MOTLAGH, S ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1187-1199
[9]   MEMBRANE GAS-CHROMATOGRAPHIC SYSTEM FOR AUTOMATED EXTRACTION AND DETERMINATION OF TRACE ORGANICS IN AQUEOUS SAMPLES [J].
MELCHER, RG ;
MORABITO, PL .
ANALYTICAL CHEMISTRY, 1990, 62 (20) :2183-2188
[10]   ISOLATION AND PRECONCENTRATION OF VOLATILE ORGANIC-COMPOUNDS FROM WATER [J].
NAMIESNIK, J ;
GORECKI, T ;
BIZIUK, M ;
TORRES, L .
ANALYTICA CHIMICA ACTA, 1990, 237 (01) :1-60