OPTIMIZATION OF THE INSTRUMENTAL PARAMETERS OF A COMBINED LIQUID CHROMATOGRAPH MASS-SPECTROMETER, COUPLED BY AN INTERFACE FOR DIRECT LIQUID INTRODUCTION .1. PERFORMANCE OF THE VACUUM EQUIPMENT

被引:33
作者
ARPINO, PJ [1 ]
GUIOCHON, G [1 ]
KRIEN, P [1 ]
DEVANT, G [1 ]
机构
[1] NERMAG INC,F-92502 RUEIL MALMAISON,FRANCE
来源
JOURNAL OF CHROMATOGRAPHY | 1979年 / 185卷 / DEC期
关键词
D O I
10.1016/S0021-9673(00)85628-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The direct introduction of liquid solutions into the vacuum space of a mass spectrometer for liquid chromatography-mass spectrometry (LC-MS) purposes requires that the vacuum pumping equipment for the apparatus evacuates very high throughputs of gas resulting from the vaporization of the solvent. Updating and testing of the vacuum equipment are then the first operations to be carried out before adapting a direct liquid introduction (DLI) interface to a chemical ionization mass spectrometer. The requirements of a sucessful DLI/LC-MS experiment are examined. The need for a cryopump with a large active surface area chilled at 80°K is emphasized. © 1979.
引用
收藏
页码:529 / 547
页数:19
相关论文
共 16 条
[1]   LIQUID CHROMATOGRAPHY MASS SPECTROMETRY .2. CONTINUOUS MONITORING [J].
ARPINO, P ;
BALDWIN, MA ;
MCLAFFERTY, FW .
BIOMEDICAL MASS SPECTROMETRY, 1974, 1 (01) :80-82
[2]  
ARPINO PJ, 1979, ANAL CHEM, V51, pA682
[3]   LIQUID CHROMATOGRAPHY MASS SPECTROMETRY SYSTEM PROVIDING CONTINUOUS MONITORING WITH NANOGRAM SENSITIVITY [J].
ARPINO, PJ ;
DAWKINS, BG ;
MCLAFFER.FW .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1974, 12 (10) :574-578
[4]  
ARPINO PJ, 1978, 26TH ANN C MASS SPEC, P426
[5]   CRYOPUMPING IN TRANSITION AND CONTINUUM PRESSURE REGIONS [J].
DAVEY, G .
VACUUM, 1976, 26 (01) :17-22
[6]   MASS-SPECTROMETRIC STUDIES OF PEPTIDES .9. POLYPEPTIDE SEQUENCING BY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
DAWKINS, BG ;
ARPINO, PJ ;
MCLAFFERTY, FW .
BIOMEDICAL MASS SPECTROMETRY, 1978, 5 (01) :1-6
[7]  
Dawson P.H., 1976, QUADRUPOLE MASS SPEC
[8]   INTRODUCTION TO CRYOPUMP DESIGN [J].
HANDS, BA .
VACUUM, 1976, 26 (01) :11-16
[10]  
HENION JD, 1978, 26TH ANN C MASS SPEC, P420