QUANTIFICATION OF MONOCLONAL-ANTIBODIES IN COMPLEX-MIXTURES BY PROTEIN-G HIGH-PERFORMANCE LIQUID AFFINITY-CHROMATOGRAPHY

被引:10
作者
BLANK, GS
VETTERLEIN, D
机构
[1] Department of Recovery Process R and D, Genentech, Inc., South San Francisco, CA 94080
关键词
D O I
10.1016/0003-2697(90)90201-J
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
High-performance liquid affinity chromatography (HPLAC) utilizing Protein G as a ligand has been evaluated for rapid quantification of monoclonal antibodies (MAbs) in various solutions. The results obtained by HPLAC agreed to within 10% of a standard enzyme-linked immunospecific assay (ELISA). A standard curve was prepared by injection of known amounts of a purified murine IgG1 with the elution peak area analyzed by computer integration software. Accuracy of quantification was independent of the injection volume, solution composition, or mouse IgG subclass. A method is described for using Protein G HPLAC to determine murine IgG levels in various complex mixtures within 15 min, compared to the ELISA which required 5 h. © 1990.
引用
收藏
页码:317 / 320
页数:4
相关论文
共 15 条
[1]  
AKERSTROM B, 1985, J IMMUNOL, V135, P2589
[2]  
AKERSTROM B, 1986, J BIOL CHEM, V261, P240
[3]   RAPID CHROMATOGRAPHIC MONITORING OF BIOPROCESSES [J].
CHASE, HA .
BIOSENSORS, 1986, 2 (05) :269-286
[4]  
DUFFY SA, 1989, BIOPHARM, V2, P34
[5]   ISOLATION OF PURE IGG1, IGG2A AND IGG2B IMMUNOGLOBULINS FROM MOUSE SERUM USING PROTEIN A-SEPHAROSE [J].
EY, PL ;
PROWSE, SJ ;
JENKIN, CR .
IMMUNOCHEMISTRY, 1978, 15 (07) :429-436
[7]  
HAMMEN R F, 1988, Biochromatography, V3, P54
[8]  
KRONVALL G, 1973, J IMMUNOL, V111, P1401
[9]   CLEAVAGE OF STRUCTURAL PROTEINS DURING ASSEMBLY OF HEAD OF BACTERIOPHAGE-T4 [J].
LAEMMLI, UK .
NATURE, 1970, 227 (5259) :680-+
[10]  
LARSSON PO, 1984, METHOD ENZYMOL, V104, P212