THE USE OF IMPLANTABLE SENSORS FOR THE CONTROL OF PACEMAKER MEDIATED TACHYCARDIAS - A COMPARATIVE-EVALUATION BETWEEN MINUTE VENTILATION SENSING AND ACCELERATION SENSING DUAL CHAMBER RATE ADAPTIVE PACEMAKERS

被引:16
作者
LAU, CP
TAI, YT
FONG, PC
LI, JPS
CHUNG, FLW
SONG, S
机构
[1] Division of Cardiology, Department of Medicine, Queen Mary Hospital
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1992年 / 15卷 / 01期
关键词
DUAL CHAMBER RATE ADAPTIVE PACING; SENSORS; PACEMAKER MEDIATED TACHYCARDIAS; PACEMAKER INTERFERENCE;
D O I
10.1111/j.1540-8159.1992.tb02899.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The role of implantable sensors to control pacemaker mediated tachycardias was investigated in 16 patients with two different dual chamber rate adaptive (DDDR) pacemakers, which sensed either minute ventilation (DDDR-Meta, nine patients) or body acceleration (Relay, seven patients). Successive atrial sensed events beyond a programmable rate occurring in the absence of detection of exercise by the sensors were considered to represent retrograde conduction or atrial arrhythmias, and the pacemakers responded by either a mode shift from DDDR to ventricular rate adaptive (VVIR) pacing (DDDR-Meta) or by tracking at an interim rate, the so-called conditional ventricular tracking limit (CVTL, Relay). In the unipolar atrial sensing mode, myopotential sensing (MPI) and external chest wall stimulations (CWS) at 250 beats/min were induced to be preferentially sensed by the atrial channel to simulate the conditions of atrial arrhythmias. In the DDD mode, these maneuvers resulted in ventricular responses of 88 +/- 3 beats/min and 110 +/- 3 beats/min for MPI and CWS, respectively. The pacing rate was significantly reduced in the DDDR mode with the sensors correctly detecting and responding to the sensed abnormal atrial signals (68 +/- 5 beats/min during MPI and 71 +/- 5 beats/min during CWS, P < 0.005 compared with the corresponding DDD rate). One patient with a Relay pacemaker developed spontaneous atrial flutter and the ventricular tracking responses were 140 and 85 beats/min in the DDD and DDDR pacing modes, respectively. Thus MPI and CWS are useful bedside testing methods to assess pacemaker response during atrial arrhythmias. The use of implantable sensors to judge the appropriateness of atrial rate is a new approach to the management of pacemaker mediated tachycardias.
引用
收藏
页码:34 / 44
页数:11
相关论文
共 23 条
  • [1] THE BASIS FOR ACTIVITY CONTROLLED RATE VARIABLE CARDIAC-PACEMAKERS - AN ANALYSIS OF MECHANICAL FORCES ON THE HUMAN-BODY INDUCED BY EXERCISE AND ENVIRONMENT
    ALT, E
    MATULA, M
    THERES, H
    HEINZ, M
    BAKER, R
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (10): : 1667 - 1680
  • [2] BAROLD SS, 1985, MODERN CARDIAC PACIN, P439
  • [3] BAROLD SS, 1985, CARDIAC PACING ELECT, P608
  • [4] BELOT PH, 1985, MODERN CARDIAC PACIN, P625
  • [5] DDD PACEMAKERS MAXIMIZE HEMODYNAMIC BENEFITS AND MINIMIZE COMPLICATIONS FOR MOST PATIENTS
    BYRD, CL
    SCHWARTZ, SJ
    GONZALES, M
    BYRD, CB
    CIRALDO, RJ
    SIVINA, M
    YAHR, WZ
    GREENBERG, JJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11): : 1911 - 1916
  • [6] ARRHYTHMIAS OF DUAL CHAMBER PACEMAKERS
    FURMAN, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (04): : 469 - 470
  • [7] CLINICAL PREDICTORS AND NATURAL-HISTORY OF ATRIAL-FIBRILLATION IN PATIENTS WITH DDD PACEMAKERS
    GROSS, J
    MOSER, S
    BENEDEK, ZM
    ANDREWS, C
    FURMAN, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1828 - 1831
  • [8] MYOPOTENTIAL INTERFERENCE WITH DDD PACEMAKERS - ENDOCARDIAL ELECTROGRAPHIC TELEMETRY IN THE DIAGNOSIS OF PACEMAKER-RELATED ARRHYTHMIAS
    HALPERIN, JL
    CAMUNAS, JL
    STERN, EH
    ROTHLAUF, EB
    KUPERSMITH, J
    ESTIOKO, MR
    MACE, RC
    STEINMETZ, MY
    TEICHHOLZ, LE
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1984, 54 (01) : 97 - 102
  • [9] ACTIVITY-SENSING, RATE-RESPONSIVE PACING - IMPROVEMENT IN MYOCARDIAL PERFORMANCE WITH EXERCISE
    HUMEN, DP
    KOSTUK, WJ
    KLEIN, GJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (01): : 52 - 59
  • [10] SUPERIOR CARDIAC HEMODYNAMICS OF ATRIOVENTRICULAR SYNCHRONY OVER RATE RESPONSIVE PACING AT SUBMAXIMAL EXERCISE - OBSERVATIONS IN ACTIVITY SENSING DDDR PACEMAKERS
    LAU, CP
    WONG, CK
    LEUNG, WH
    LIU, WX
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1832 - 1837