EXPERIMENTAL EVALUATION OF A NEW COMPOSITE MESH WITH THE SELECTIVE PROPERTY OF INCORPORATION TO THE ABDOMINAL-WALL WITHOUT ADHERING TO THE INTESTINES

被引:99
作者
AMID, PK
SHULMAN, AG
LICHTENSTEIN, IL
SOSTRIN, S
YOUNG, J
HAKAKHA, M
机构
[1] HARBOR UCLA,DEPT SURG,LOS ANGELES,CA
[2] HARBOR UCLA,INST RES & EDUC,LOS ANGELES,CA
[3] CEDARS SINAI MED CTR,DEPT SURG,LOS ANGELES,CA 90048
[4] CEDARS SINAI MED CTR,DEPT COMPARAT MED,LOS ANGELES,CA
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1994年 / 28卷 / 03期
关键词
D O I
10.1002/jbm.820280312
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This preliminary study examined the possibility of preventing intestinal adhesions to biomaterials while preserving their incorporation with the abdominal wall. White New Zealand rabbits received intraperitoneal implants of different biomaterials for repair of defects created on the abdominal wall. The following biomaterials were used: polypropylene, polyester, expanded polytetraflueroethylene, polypropylene mesh/polypropylene sheeting (polypropylene mesh covered with polypropylene sheeting on the peritoneal side), polypropylene/silastic, polypropylene/polyglactin, polypropylene/polyglycolic acid, and polypropylene/fibrin. All biomaterials evaluated caused adhesions to the intestines except for polypropylene mesh/polypropylene sheeting and polypropylene mesh/silastic composites. Because adhesion of the intestine to the biomaterial is the first stage of biomaterial-related intestinal fistula, its prevention is logical for the elimination of this complication. Composites with the selective property of adhering to the abdominal wall, yet sparing the viscera, would facilitate thoracic and abdominal wall surgeries, as well as intraperitoneal laparoscopic hernioplasties. (C) 1994 John Wiley and Sons, Inc.
引用
收藏
页码:373 / 375
页数:3
相关论文
共 12 条
[1]  
GEORGE CD, 1986, ANN ROY COLL SURG, V68, P185
[2]   INCISIONAL HERNIA [J].
HORTON, RE ;
SMITH, PC .
PROCEEDINGS OF THE ROYAL SOCIETY OF MEDICINE-LONDON, 1969, 62 (05) :513-&
[3]   FECAL FISTULA - A LATE COMPLICATION OF MARLEX MESH REPAIR [J].
KAUFMAN, Z ;
ENGELBERG, M ;
ZAGER, M .
DISEASES OF THE COLON & RECTUM, 1981, 24 (07) :543-544
[4]  
Law N., 1988, CLIN MATER, V3, P95
[5]   ABDOMINAL WOUND CLOSURE - TRIAL OF NYLON, POLYGLYCOLIC ACID AND STEEL SUTURES [J].
LEAPER, DJ ;
POLLOCK, AV ;
EVANS, M .
BRITISH JOURNAL OF SURGERY, 1977, 64 (08) :603-606
[6]  
LEWIS RT, 1962, ARCH SURG-CHICAGO, P325
[7]  
LOURY JN, 1993, TRAITEMENT EVENTRATI, P2116
[8]   INCISIONAL HERNIA - A 10 YEAR PROSPECTIVE-STUDY OF INCIDENCE AND ATTITUDES [J].
MUDGE, M ;
HUGHES, LE .
BRITISH JOURNAL OF SURGERY, 1985, 72 (01) :70-71
[9]  
REISFELD D, 1989, SURG ROUNDS, V12, P69
[10]  
SCHNEIDER R, 1979, AM SURGEON, V45, P337