TOTAL-BODY SUPERFICIAL ELECTRON-BEAM THERAPY USING A MULTIPLE-FIELD PENDULUM-ARC TECHNIQUE

被引:20
作者
SEWCHAND, W
KHAN, FM
WILLIAMSON, J
机构
[1] UNIV MINNESOTA HOSP,DEPT THERAPEUT RADIOL,MINNEAPOLIS,MN 55455
[2] HARSHAW CHEM CO,SOLON,OH
关键词
D O I
10.1148/130.2.493
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A technique using pendulum-arc rotation is presented for electron-beam treatment of generalized superficial malignancies. The technique consists of six arcing fields symmetrically dispersed around the body surface for circumferential coverage. The arc angle is selected to scan the height of the body fully. Beam uniformity within 10% over a height of 180 cm is achieved at a treatment distance of 385 cm. Randophantom dosimetry, using an 8 MeV electron beam degraded by 3/8 in Plexiglas, indicates a surface dose uniformly within ±7% over most of the body surface. Underdosages occur at regions obstructed by adjacent body parts.
引用
收藏
页码:493 / 498
页数:6
相关论文
共 12 条
[1]  
BJARNGARD BE, 1977, INT J RADIAT ONCOL, V2, P319, DOI 10.1016/0360-3016(77)90090-6
[2]   DOSIMETRY FOR TOTAL-BODY ELECTRON-BEAM THERAPY IN TREATMENT OF MYCOSIS-FUNGOIDES [J].
EDELSTEIN, GR ;
CLARK, T ;
HOLT, JG .
RADIOLOGY, 1973, 108 (03) :691-694
[3]  
Feldman A, 1974, Med Phys, V1, P74, DOI 10.1118/1.1637283
[4]   TOTAL-SKIN ELECTRON TREATMENT OF MYCOSIS FUNGOIDES [J].
FUKS, Z ;
BAGSHAW, MA .
RADIOLOGY, 1971, 100 (01) :145-&
[5]  
HELLER E H, 1972, Australasian Journal of Dermatology, V13, P11, DOI 10.1111/j.1440-0960.1972.tb00477.x
[6]   A TECHNIQUE FOR LARGE-FIELD, SUPERFICIAL ELECTRON THERAPY [J].
KARZMARK, CJ ;
LOEVINGER, R ;
STEELE, RE ;
WEISSBLUTH, M .
RADIOLOGY, 1960, 74 (04) :633-644
[7]   FIELD SHAPING IN ELECTRON-BEAM THERAPY [J].
KHAN, FM ;
MOORE, VC ;
LEVITT, SH .
BRITISH JOURNAL OF RADIOLOGY, 1976, 49 (586) :883-886
[8]  
MARKUS B, 1964, Strahlentherapie, V123, P350
[9]   PATIENT DOSIMETRY IN ELECTRON TREATMENT OF LARGE SUPERFICIAL LESIONS [J].
PAGE, V ;
GARDNER, A ;
KARZMARK, CJ .
RADIOLOGY, 1970, 94 (03) :635-&
[10]   BACKSCATTERING FROM METALS IN SUPERFICIAL THERAPY WITH HIGH-ENERGY ELECTRONS [J].
SAUNDERS, JE ;
PETERS, VG .
BRITISH JOURNAL OF RADIOLOGY, 1974, 47 (560) :467-470