INAPPROPRIATE DISCHARGES BY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DURING POSTOPERATIVE TESTING - IMPLICATIONS FOR INTRAOPERATIVE ASSESSMENT

被引:4
作者
SULKE, N
HOLT, P
BOSTOCK, J
YATES, A
SOWTON, E
机构
[1] Department of Cardiology., Guy's Hospital, London
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 09期
关键词
ICD telemetry interaction; Intraoperative testing;
D O I
10.1111/j.1540-8159.1990.tb02169.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inappropriate shocks were delivered to a patient while in sinus rhythm by an implantable Cardioverter defibrillator (ICD) during routine prehospitai discharge testing. This was induced by the standard programmer when the “read” telemetry sequence was initiated. The ICD was removed and found to suffer from electrical artifact that was sensed as ventricular tachycardia during telemetry. To avoid inadvertent telemetry‐induced shocks during routine testing, all ICDs should be interrogated, using a standard programmer, intraoperatively, with the unit in “defibrillation on” mode. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:1123 / 1126
页数:4
相关论文
共 6 条
[1]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EVALUATING SUSPECTED INAPPROPRIATE SHOCKS [J].
CHAPMAN, PD ;
TROUP, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 7 (05) :1075-1078
[2]   THE USE AND INTERACTION OF PERMANENT PACEMAKERS AND THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
COHEN, AI ;
WISH, MH ;
FLETCHER, RD ;
MILLER, FC ;
MCCORMICK, D ;
SHUCK, J ;
SHAPIRA, N ;
DELNEGRO, AA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (06) :704-711
[3]   CLINICAL-EXPERIENCE, COMPLICATIONS, AND SURVIVAL IN 70 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
ECHT, DS ;
ARMSTRONG, K ;
SCHMIDT, P ;
OYER, PE ;
STINSON, EB ;
WINKLE, RA .
CIRCULATION, 1985, 71 (02) :289-296
[4]   AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - INADVERTENT DISCHARGES DURING PERMANENT PACEMAKER MAGNET TESTS [J].
KIM, SG ;
FURMAN, S ;
MATOS, JA ;
WASPE, LE ;
BRODMAN, R ;
FISHER, JD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03) :579-582
[5]  
MIROWSKI M, 1985, J AM COLL CARDIOL, V6, P460
[6]   TREATMENT OF MALIGNANT VENTRICULAR ARRHYTHMIAS WITH ENDOCARDIAL RESECTION AND IMPLANTATION OF THE AUTOMATIC CARDIOVERTER-DEFIBRILLATOR [J].
PLATIA, EV ;
GRIFFITH, LSC ;
WATKINS, L ;
MOWER, MM ;
GUARNIERI, T ;
MIROWSKI, M ;
REID, PR .
NEW ENGLAND JOURNAL OF MEDICINE, 1986, 314 (04) :213-216