THE RANCHO EMG ANALYZER - A COMPUTERIZED SYSTEM FOR GAIT ANALYSIS

被引:27
作者
PERRY, J [1 ]
BONTRAGER, EL [1 ]
BOGEY, RA [1 ]
GRONLEY, JK [1 ]
BARNES, LA [1 ]
机构
[1] RANCHO LOS AMIGOS MED CTR,DOWNEY,CA 90242
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1993年 / 15卷 / 06期
关键词
EMG; GAIT; COMPUTER ANALYSIS; FOOTSWITCHES;
D O I
10.1016/0141-5425(93)90064-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes a computer system which accurately defines the EMG patterns of the lower extremities during gait. Footswitches are used to identify the temporal relationships and determine the phases of the gait cycle. Fine wire electrodes, inserted in the desired muscles of the patient being tested, provide EMG signals for comparison with a normal database. The system is also usable with surface electrodes when an appropriate normal database for surface electrodes is incorPorated. Descriptive qualifiers (such as 'premature onset, 'delayed cessation, 'no clinically significant EMG, 'continuous activity' etc.) are used to produce a clinically relevant printed (textual) report. The intensity filtered average (IFA) of the EMG is shown graphically with the representative profile of each stride. The IFAs for all muscles tested can be plotted together (up to six on a page) and the graphic representation of the 'raw' EMG can be produced. The methods of generating the normal database by creating time-adjusted mean profiles (TAMP) are enumerated. The clinical use of the system is discussed. A detailed analysis of 31 of the most recent patient tests for which the system was used provides an indication of its accuracy. For 86% of the 428 muscle tests examined, the EMG analyser was considered to have given the correct result as compared with a visual analysis of the raw EMG record by a trained expert. Recommendations for the use and future improvements of the EMG analyser are made.
引用
收藏
页码:487 / 496
页数:10
相关论文
共 31 条
[1]  
ANTONELLI D, 1977, T ORTHOP RES SOC, V2, P182
[2]  
BARTO PS, 1984, DEV MED CHILD NEUROL, V26, P88
[3]   A NEW BIPOLAR ELECTRODE FOR ELECTROMYOGRAPHY [J].
BASMAJIAN, JV ;
STECKO, G .
JOURNAL OF APPLIED PHYSIOLOGY, 1962, 17 (05) :849-849
[4]   PATTERN-RECOGNITION OF MULTIPLE EMG SIGNALS APPLIED TO DESCRIPTION OF HUMAN GAIT [J].
BEKEY, GA ;
CHANG, CW ;
PERRY, J ;
HOFFER, MM .
PROCEEDINGS OF THE IEEE, 1977, 65 (05) :674-681
[5]  
BOGEY RA, 1992, ARCH PHYS MED REHAB, V73, P835
[6]  
Bontrager E. L., 1990, Fourteenth Annual Symposium on Computer Applications in Medical Care. Standards in Medical Informatics. A Conference of the American Medical Informatics Association, P625
[7]  
CLYDE K, 1975, THESIS U SO CALIFORN
[8]   SURFACE MYOELECTRIC SIGNAL CROSS-TALK AMONG MUSCLES OF THE LEG [J].
DE LUCA, CJ ;
MERLETTI, R .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1988, 69 (06) :568-575
[9]  
EBERHART HD, 1968, HUMAN LIMBS THEIR SU, P437
[10]  
GAGE JR, 1983, B HOSP JOINT DIS ORT, V43, P147