DETERMINATION OF VOLATILE MERCURY-COMPOUNDS IN AIR WITH THE COLEMAN MERCURY ANALYZER SYSTEM

被引:44
作者
DUMAREY, R
HEINDRYCKX, R
DAMS, R
HOSTE, J
机构
关键词
D O I
10.1016/S0003-2670(01)93206-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple but very selective cold-vapour atomic absorption system is described for the determination of volatile mercury compounds at very low levels in ambient air. Three different absorbers are compared: activated charcoal, silver-coated sea sand and gold-coated sea sand. To eliminate interferences, a two-step desorption unit is used. After thermal desorption, the mercury is measured by using a modified MAS-50 spectrophotometer. The effects of flow rate and desorption temperature are discussed. The detection limit is 0.1 ng. Above 1 ng, the reproducibility is about 1%. Calibration is done by injection of elemental mercury vapour. The method with gold-coated sand absorbers is most satisfactory and is suitable for the analysis of ambient outdoor and indoor air. All likely volatile mercury compounds are absorbed, and a wide range of mercury concentrations can be determined. In routine application, one analysis takes about 3 min. © 1979.