A MODEL FOR THE EFFECT OF INHALATION DEVICE FLOW RESISTANCE ON THE PEAK INSPIRATORY FLOW-RATE AND ITS APPLICATION IN PHARMACEUTICAL TESTING

被引:7
作者
OLSSON, B
ASKING, L
机构
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 1994年 / 7卷 / 02期
关键词
POWDER INHALER; DRUG DELIVERY; INSPIRATORY FLOW; FLOW RATE; INSPIRATORY FORCE; AIR FLOW RESISTANCE; ASTHMA; PARTICLE SIZE DISTRIBUTION; RESPIRABLE DOSE;
D O I
10.1089/jam.1994.7.201
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
An empirical relation between the peak inspiratory flow through a device and the flow resistance of the device has been developed using results from healthy subjects and verified to a limited extent using data from asthmatic patients. The empirical relation can be used to predict the change in peak inspiratory flow resulting from a change in flow resistance. Matching flow rates, representing equal inspiratory force, may thus be calculated for different inhalation devices. The fine particle dose generated by a variety of commercial powder inhalers was tested in-vitro at flow rates representing, for each device, a weak, a medium and a forceful inspiration. It was found that both the level of the fine particle dose and its flow rate dependence varied for the products.