ASCORBIC-ACID DETERMINATION IN BIOLOGICAL-FLUIDS USING ASCORBATE OXIDASE IMMOBILIZED ON ALKYLAMINE GLASS-BEADS IN A FLOW-INJECTION POTENTIOMETRIC SYSTEM

被引:9
作者
MARQUES, IDHC
MARQUES, ETA
SILVA, AC
LEDINGHAM, WM
MELO, EHM
DASILVA, VL
LIMA, JL
机构
[1] UNIV FED PERNAMBUCO,LIKA,DEPT BIOQUIM,BR-50730 RECIFE,PE,BRAZIL
[2] UNIV FED PERNAMBUCO,LIKA,DEPT ENGN QUIM,BR-50730 RECIFE,PE,BRAZIL
关键词
FLOW INJECTION ANALYSIS (FIA); ASCORBIC ACID; OXYGEN ELECTRODE; GLASS BEADS; CUCURBITA MAXIMA;
D O I
10.1007/BF02921853
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
A flow injection method was developed aimed at the determination of ascorbic acid in biological fluids, particularly fruit juices. The enzyme ascorbic oxidoreductase (EC 1.10.3.3), extracted from Cucurbita maxima, was immobilized onto alkylamine glass beads using glutaraldehyde as a bifunctional agent. The ascorbic acid concentration was related to oxygen saturation. Fall in oxygen concentration, as a result of ascorbic acid oxidation, was detected by a low cost, homemade oxygen electrode. The calibration graph was linear over the range 0.05 to 3.00 mM (RSD 1%), the maximum number of samples that could be analysed was 90/h. The immobilized enzyme retained its initial activity for 2 mo with more than 600 assays.
引用
收藏
页码:81 / 89
页数:9
相关论文
共 16 条
[1]   ASCORBATE OXIDASE FROM CUCURBITA-MAXIMA [J].
CARVALHO, LB ;
LIMA, CJ ;
MEDEIROS, PH .
PHYTOCHEMISTRY, 1981, 20 (10) :2423-2424
[2]   APPLICATION OF SQUARE-WAVE VOLTAMMETRY FOR THE DETERMINATION OF ASCORBIC-ACID IN SOFT DRINKS AND FRUIT JUICES USING A FLOW-INJECTION SYSTEM [J].
FUNG, YS ;
MO, SY .
ANALYTICA CHIMICA ACTA, 1992, 261 (1-2) :375-380
[3]   DETERMINATION OF L-ASCORBIC-ACID IN FRUIT AND VEGETABLE JUICES BY FLOW-INJECTION WITH IMMOBILIZED ASCORBATE OXIDASE [J].
GREENWAY, GM ;
ONGOMO, P .
ANALYST, 1990, 115 (10) :1297-1299
[4]   DETERMINATION OF ASCORBIC-ACID USING AN ORGANIC CONDUCTING SALT ELECTRODE [J].
KORELL, U ;
LENNOX, RB .
ANALYTICAL CHEMISTRY, 1992, 64 (02) :147-151
[5]  
LAZARO F, 1987, ANALUSIS, V15, P183
[6]  
LOWRY OH, 1951, J BIOL CHEM, V193, P265
[7]   ASCORBIC-ACID BIOSENSOR USING ASCORBATE OXIDASE IMMOBILIZED ON ALKYLAMINE GLASS-BEADS [J].
MARQUES, ETA ;
LIMA, JL .
APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 1992, 32 :73-78
[8]   A LOW-COST OXYGEN METER [J].
MARQUES, ETH ;
NETTO, FACS ;
FILHO, JLL .
BIOCHEMICAL EDUCATION, 1991, 19 (02) :77-78
[9]   ANALYSIS OF ASCORBIC-ACID IN HUMAN-PLASMA AND URINE BY HPLC WITH ELECTROCHEMICAL DETECTION [J].
MASON, WD ;
AMICK, EN ;
HEFT, W .
ANALYTICAL LETTERS PART B-CLINICAL AND BIOCHEMICAL ANALYSIS, 1980, 13 (10) :817-824
[10]  
MCCORNICK DB, 1979, METHODS ENZYMOLOGY D, V62, P3