STANDARDIZATION OF ABSOLUTE CD4(+) LYMPHOCYTE COUNTS ACROSS LABORATORIES - AN EVALUATION OF THE ORTHO CYTORONABSOLUTE FLOW-CYTOMETRY SYSTEM ON NORMAL DONORS

被引:49
作者
CONNELLY, MC
KNIGHT, M
GIORGI, JV
KAGAN, J
LANDAY, AL
PARKER, JW
PAGE, E
SPINO, C
WILKENING, C
MERCOLINO, TJ
机构
[1] UNIV CALIF LOS ANGELES,SCH MED,DEPT MED,LOS ANGELES,CA 90024
[2] NIAID,DIV AIDS,ROCKVILLE,MD
[3] RUSH PRESBYTERIAN ST LUKES MED CTR,CHICAGO,IL 60612
[4] UNIV SO CALIF,SCH MED,DEPT PATHOL,LOS ANGELES,CA 90033
[5] HARVARD UNIV,SCH PUBL HLTH,BOSTON,MA 02115
来源
CYTOMETRY | 1995年 / 22卷 / 03期
关键词
CD4(+); LYMPHOCYTE COUNTS; FLOW CYTOMETRY;
D O I
10.1002/cyto.990220307
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The Ortho CytoronAbsolute is a flow cytometer designed to provide direct absolute counts of lymphocytes and their subsets from a single instrument, This study was designed to determine the performance of four geographically separated CytoronAbsolute instruments using 24-h-old, shipped, whole blood samples and to compare the results obtained on the OytoronAbsolute to those obtained using combinations of hematology instruments and other flow cytometers, The absolute count feature of the CytoronAbsolutes located at the four sites were cross calibrated and gave across-site coefficients of variation (CVs) of <4.0% for absolute count and 8.2% for absolute lymphocyte count, The calibration was stable for at least 2 months, Absolute lymphocyte counts and lymphocyte percentage immunophenotypes were determined on blood from 50 healthy human immunodeficiency virus (HIV)-seronegative donors, There were no significant site-to-site differences (each P >.05) in CD3(+)/CD4(+) absolute lymphocyte counts determined on the CytoronAbsolute. In contrast, there was a significant site-to-site difference (P <.001) between sites 2 and 3 and sites 3 and 4 in the absolute CD3(+)/CD4(+) lymphocyte counts determined via the conventional method of combining a flow cytometry-derived percentage with a hematology instrument-derived lymphocyte count, There was no significant difference (P =.388) in CD3(+)/CD4(+) lymphocyte percent determinations between the CytoronAbsolute and the FACScan or Profile II flow cytometers used in this study, These results demonstrate that different operators can cross calibrate CytoronAbsolutes for absolute CD3(+)/CD4(+) lymphocyte subset determinations, even over large geographic distances. (C) 1995 Wiley-Liss, Inc.
引用
收藏
页码:200 / 210
页数:11
相关论文
共 24 条
[1]   CONSISTENCY OF ROUTINE MEASUREMENTS OF CD4+, CD8+ PERIPHERAL-BLOOD LYMPHOCYTES [J].
ABOULKER, JP ;
AUTRAN, B ;
BELDJORD, K ;
TOURAINE, F ;
DEBRE, P .
JOURNAL OF IMMUNOLOGICAL METHODS, 1992, 154 (02) :155-161
[2]  
ADELMAN LM, 1993, J ACQ IMMUN DEF SYND, V6, P144
[3]   A COMPARISON OF THE COULTER STKS, COULTER S+IV, AND MANUAL ANALYSIS OF WHITE BLOOD-CELL DIFFERENTIAL COUNTS IN A HUMAN IMMUNODEFICIENCY VIRUS-INFECTED POPULATION [J].
COHEN, AJ ;
PEERSCHKE, EIB ;
STEIGBIGEL, RT .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1993, 100 (06) :611-617
[4]   THE PROGNOSTIC VALUE OF CELLULAR AND SEROLOGIC MARKERS IN INFECTION WITH HUMAN IMMUNODEFICIENCY VIRUS TYPE-1 [J].
FAHEY, JL ;
TAYLOR, JMG ;
DETELS, R ;
HOFMANN, B ;
MELMED, R ;
NISHANIAN, P ;
GIORGI, JV .
NEW ENGLAND JOURNAL OF MEDICINE, 1990, 322 (03) :166-172
[5]   IMMUNOLOGICAL AND SEROLOGICAL MARKERS PREDICTIVE OF PROGRESSION TO AIDS IN A COHORT OF HIV-INFECTED DRUG-USERS [J].
FERNANDEZCRUZ, E ;
DESCO, M ;
MONTES, MG ;
LONGO, N ;
GONZALEZ, B ;
ZABAY, JM .
AIDS, 1990, 4 (10) :987-994
[6]   QUALITY-CONTROL IN THE FLOW CYTOMETRIC MEASUREMENT OF LYMPHOCYTE-T SUBSETS - THE MULTICENTER AIDS COHORT STUDY EXPERIENCE [J].
GIORGI, JV ;
CHENG, HL ;
MARGOLICK, JB ;
BAUER, KD ;
FERBAS, J ;
WAXDAL, M ;
SCHMID, I ;
HULTIN, LE ;
JACKSON, AL ;
PARK, L ;
TAYLOR, JMG .
CLINICAL IMMUNOLOGY AND IMMUNOPATHOLOGY, 1990, 55 (02) :173-186
[7]   EFFECT OF T4 COUNT AND COFACTORS ON THE INCIDENCE OF AIDS IN HOMOSEXUAL MEN INFECTED WITH HUMAN-IMMUNODEFICIENCY-VIRUS [J].
GOEDERT, JJ ;
BIGGAR, RJ ;
MELBYE, M ;
MANN, DL ;
WILSON, S ;
GAIL, MH ;
GROSSMAN, RJ ;
DIGIOIA, RA ;
SANCHEZ, WC ;
WEISS, SH ;
BLATTNER, WA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1987, 257 (03) :331-334
[8]   PREVENTIVE HEALTH-CARE FOR ADULTS WITH HIV-INFECTION [J].
JEWETT, JF ;
HECHT, FM .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1993, 269 (09) :1144-1153
[9]   PRECISION AND ACCURACY OF ABSOLUTE LYMPHOCYTE COUNTS [J].
KOEPKE, JA ;
LANDAY, AL .
CLINICAL IMMUNOLOGY AND IMMUNOPATHOLOGY, 1989, 52 (01) :19-27
[10]  
LANG W, 1989, J ACQ IMMUN DEF SYND, V2, P63