ADAPTIVE RADIOFREQUENCY HYPERTHERMIA-PHASED ARRAY SYSTEM FOR IMPROVED CANCER-THERAPY - PHANTOM TARGET MEASUREMENTS

被引:13
作者
FENN, AJ [1 ]
KING, GA [1 ]
机构
[1] SUNY,HLTH SCI CTR,DIV RADIAT ONCOL,NEW YORK,NY
关键词
HYPERTHERMIA; HOT SPOTS; ADAPTIVE PHASED ARRAY; DEEP-SEATED CANCER; PHANTOM MEASUREMENTS;
D O I
10.3109/02656739409009343
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A computer-controlled adaptive radio-frequency hyperthermia system for improved therapeutic tumour heating is experimentally investigated. Adaptive array feedback techniques are used to modify the electric-field and temperature distribution in hyperthermia experiments with homogenous and heterogeneous phantom targets. A commercial hyperthermia phased-array antenna system at the SUNY Health Science Center in Syracuse, New York, has been modified to implement adaptive nulling and adaptive focusing algorithms. The hyperthermia system is the BSD Medical Corporation Model BSD-2000 with Sigma-60 annular phased-array antenna applicator. The transmit phased array system is made adaptive by software modifications which invoke a gradient-search feedback algorithm. The gradient-search algorithm implements the method of steepest descent for adaptive nulling (power minimization) and the method of steepest ascent for adaptive focusing (power maximization). The feedback signals are provided by electric-field short-dipole probe antennas. With an adaptive hyperthermia array using real-time measured data, it may be possible to maximize the applied electric field at a tumour position in a complex scattering target body and simultaneously minimize or reduce the electric field at target positions where undesired high-temperature regions (hot spots) occur. The measured phantom-target data indicate that adaptive nulling can reduce the electric field at one or more target positions while simultaneously focusing the electric field at a deep-seated position within the target.
引用
收藏
页码:189 / 208
页数:20
相关论文
共 58 条
[31]   OPTIMAL TEMPERATURE CONTROL WITH PHASED-ARRAY HYPERTHERMIA SYSTEM [J].
KNUDSEN, M ;
HARTMANN, U .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) :597-603
[32]  
LIN J, 1986, IEEE T MICROWAVE THE, V34
[33]   GAIN OPTIMIZATION OF A NEAR-FIELD FOCUSING ARRAY FOR HYPERTHERMIA APPLICATIONS [J].
LOANE, JT ;
LEE, SW .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1989, 37 (10) :1629-1635
[34]   AN OPTIMAL EXCITATION METHOD IN MULTI-APPLICATOR SYSTEMS FOR FORMING A HOT ZONE INSIDE THE HUMAN-BODY [J].
MORITA, N ;
HAMASAKI, T ;
KUMAGAI, N .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) :532-538
[35]  
NGUYEN DB, 1989, HYPERTHERMIC ONCOLOG, P738
[36]   DETERMINATION OF POWER DEPOSITION PATTERNS FOR LOCALIZED HYPERTHERMIA - A TRANSIENT ANALYSIS [J].
OCHELTREE, KB ;
FRIZZELL, LA .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (03) :281-296
[37]  
OVERGAARD J, 1978, CANCER THERAPY HYPER, P49
[38]  
OVERGAARD J, 1987, 8TH P INT C RAD RES, V2, P942
[39]  
PEREZ CA, 1987, PRINCIPLES PRACTICE
[40]   OPTIMAL SOURCE DISTRIBUTION FOR HYPERTHERMIA AT THE CENTER OF A SPHERE OF MUSCLE-TISSUE [J].
RAPPAPORT, CM ;
MORGENTHALER, FR .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1987, 35 (12) :1322-1327