PERFORMANCE OF A LIQUID XENON CALORIMETER CELL

被引:4
作者
BRAEM, A
GONIDEC, A
SCHINZEL, D
SEIDL, W
CLAYTON, E
HALL, G
PAYNE, R
ROE, S
SEEZ, C
STRIEBIG, J
VIRDEE, TS
COCKERILL, D
机构
[1] UNIV LONDON IMPERIAL COLL SCI TECHNOL & MED,BLACKETT LAB,LONDON SW7 2BZ,ENGLAND
[2] RUTHERFORD APPLETON LAB,DIDCOT OX11 0QX,OXON,ENGLAND
关键词
D O I
10.1016/0168-9002(93)90373-P
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A test cell has been constructed containing liquid xenon as a scintillator, read out by silicon drift photodiodes. Results are presented from experiments with this cell in a test beam, with particular emphasis on the cells optical performance and the optical characteristics of liquid xenon at the scintillation wavelength (175 nm). Comparison of test beam results with Monte Carlo studies of the cell show the light attenuation length of liquid xenon to be approximately 40 cm, and the photon yield to be 1.5 +/- 0.5 x 10(4) photons/MeV. Calculations of the optical characteristics of the photodiodes using a three layer model for the xenon-oxide-silicon interface are compared with measurements, and the quantum efficiency is shown to be high. Ways of improving the overall optical performance of the cell are shown, based upon calculations using the stratified layer model for both detectors and cell walls.