ON THE FIELD-SIZE DEPENDENCE OF RELATIVE OUTPUT FROM A LINEAR-ACCELERATOR

被引:63
作者
DUNSCOMBE, PB [1 ]
NIEMINEN, JM [1 ]
机构
[1] LAURENTIAN UNIV,DEPT PHYS,SUDBURY P3E 2C6,ONTARIO,CANADA
关键词
OUTPUT; LINAC;
D O I
10.1118/1.596799
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The radiation output in air on the central axis of a linac photon beam has been modeled as the sum of two components. These are a point source representing radiation direct from the target and a distributed source representing scatter in the flattening filter and primary collimator. By fitting only two parameters, the ratio of the two components for a 20 X 20 field and a width parameter for the distributed source this semi-empirical model describes the relative outputs of square, symmetric rectangular, and asymmetric rectangular fields with an average error of 0.25% for the field sizes studied.
引用
收藏
页码:1441 / 1444
页数:4
相关论文
共 6 条
[1]   THE EFFECT OF COLLIMATOR BACKSCATTER RADIATION ON PHOTON OUTPUT OF LINEAR ACCELERATORS [J].
HUANG, PH ;
CHU, J ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1987, 14 (02) :268-269
[2]  
KANELLITSAS C, 1991, Medical Physics (Woodbury), V18, P618
[3]   HEAD SCATTER DATA FOR SEVERAL LINEAR ACCELERATORS (4-18MV) [J].
KASE, KR ;
SVENSSON, GK .
MEDICAL PHYSICS, 1986, 13 (04) :530-532
[4]   OUTPUT FACTOR CONSTITUENTS OF A HIGH-ENERGY PHOTON-BEAM [J].
LUXTON, G ;
ASTRAHAN, MA .
MEDICAL PHYSICS, 1988, 15 (01) :88-91
[5]   ENERGY AND ANGULAR-DISTRIBUTIONS OF PHOTONS FROM MEDICAL LINEAR ACCELERATORS [J].
MOHAN, R ;
CHUI, C ;
LIDOFSKY, L .
MEDICAL PHYSICS, 1985, 12 (05) :592-597
[6]   THE DETERMINATION OF PHANTOM AND COLLIMATOR SCATTER COMPONENTS OF THE OUTPUT OF MEGAVOLTAGE PHOTON BEAMS - MEASUREMENT OF THE COLLIMATOR SCATTER PART WITH A BEAM-COAXIAL NARROW CYLINDRICAL PHANTOM [J].
VANGASTEREN, JJM ;
HEUKELOM, S ;
VANKLEFFENS, HJ ;
VANDERLAARSE, R ;
VENSELAAR, JLM ;
WESTERMANN, CF .
RADIOTHERAPY AND ONCOLOGY, 1991, 20 (04) :250-257