LONG-TERM RESULTS AND BIOCOMPATIBILITY OF HEPARIN-SURFACE-MODIFIED INTRAOCULAR LENSES

被引:27
作者
AMON, M
MENAPACE, R
机构
关键词
BIOCOMPATIBILITY; CATARACT SURGERY; CELLULAR REACTION; HEPARIN SURFACE MODIFICATION; FIBROBLAST-LIKE CELL; FOREIGN-BODY GIANT CELL; IMPLANTATION; INTRAOCULAR LENS; INTRAOCULAR LENS DESIGN;
D O I
10.1016/S0886-3350(13)80953-4
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A prospective in vivo study of 50 intraocular lenses was done to document the long-term results and biocompatibility of heparin-surface-modified poly(methyl methacrylate) posterior chamber lenses. Continuous curvilinear capsulorhexis and phacoemulsification were performed throughout and examinations were by slit-lamp and specular microscopy. Decentrations of more than 1 mm were seen in 4% of eyes. Stress folds from the haptic traction forces were detected in 16%. A posterior capsulotomy was performed in 1 0% because of fibrosis or Elschnig pearl formation. Fine fibers were found on the lens surface in 76%. During the first postoperative days only a moderate number of fibroblast-like cells was observed. Foreign-body giant cells were seen in 8%. All cases with foreign-body giant cells had posterior synechias. Visual performance and clinical results were comparable to those of other well-approved intraocular lenses. The low percentage of cellular reaction on the lens surface suggests good biocompatibility.
引用
收藏
页码:258 / 262
页数:5
相关论文
共 22 条
  • [11] McDonnell PJ, Zarbin MA, Green WR, Posterior capsule opacification in pseudophakic eyes, Ophthalmology, 90, pp. 1533-1548, (1983)
  • [12] Ohara K, Wide-field specular microscopy, J Ophthalmic Photogr, 9, pp. 97-100, (1986)
  • [13] Greite JH, Kammann JP, Tsinopoulos I, Kreiner CF, Die ST-Linse-Eine Ganzkorperlinse zur spannungsfreien endokapsularen Fixation, 4. Kongress der Deutschen Gesellschaft fur Intraokularlinsen Implantation
  • [14] 6. bis 7. April 1990, Essen, pp. 5-12, (1991)
  • [15] Amon M, Menapace R, Evaluation of a one-piece poly(methyl methacrylate) intraocular lens with a 7 mm biconvex optic and a total diameter of 10 mm, J Cataract Refract Surg, 19, pp. 16-21, (1993)
  • [16] Drews RC, Kreiner C, Comparative study of the elasticity and memory of intraocular lens loops, J Cataract Refract Surg, 13, pp. 525-530, (1987)
  • [17] Skorpik, Menapace R, Grasl M, Amon M, Erfahrungen mit der Implantation von lochlosen PMMA-Hinterkammerlinsen mit 17-mm-Optik and eingesetzten Bfgeln, 3. Kongress der Deutschen Gesellschaft fur Intraokularlinsen Implantation
  • [18] 2. bis 4. Marz 1989, Wein, pp. 339-348, (1990)
  • [19] Wolter JR, Morphology of the capsule-like portion of the reactive membranes on intraocular lens implants, Graefes Arch Clin Exp Ophthalmol, 220, pp. 58-65, (1983)
  • [20] Wenzel M, Heinze M, Reim M, Membranen auf Intraokularlinsen-Gewebsdifl'erenzierungen in vivo, Fortschr Ophthalmol, 85, pp. 227-279, (1988)