KINETIC-ANALYSIS OF MONOCLONAL ANTIBODY-ANTIGEN INTERACTIONS WITH A NEW BIOSENSOR BASED ANALYTICAL SYSTEM

被引:986
作者
KARLSSON, R
MICHAELSSON, A
MATTSSON, L
机构
[1] Pharmacia Biosensor AB
关键词
MONOCLONAL ANTIBODY; KINETICS; ASSOCIATION RATE CONSTANT; AFFINITY CONSTANT;
D O I
10.1016/0022-1759(91)90331-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An automated biosensor system for measuring molecular interactions has been used to study the kinetics of monoclonal antibody-antigen reactions. The system combines a microfluidic unit in contact with a sensor surface for surface plasmon resonance detection. The specificity of the surface is determined by the operator. Antibody or antigen is immobilised in a dextran matrix attached to the sensor surface. The interaction of matrix bound antibody or antigen with the corresponding partner in solution is monitored in real time. None of the interacting molecules needs to be labelled and it is not necessary to determine the concentration of the matrix bound component in advance. Two systems were studied: matrix bound monoclonal antibodies (MAbs) interacting with HIV-1 core protein p24 and immobilised aminotheophylline reacting with MAbs. Control of the amount of immobilised ligand and reusable sensor surfaces permits the comparison of different MAbs reacting with antigen under almost identical conditions. Differences in affinity and reaction rates are immediately apparent. The calculated association rate constants for p24 MAbs ranged from 3 x 10(4) - 7.4 x 10(5) M-1 s-1 and for theophylline MAbs association rate constants as high as 1 x 10(6) M-1 s-1 were encountered. The calculated dissociation rate constants were in the region 2 x 10(-4) s-1 to 2 x 10(-2) s-1.
引用
收藏
页码:229 / 240
页数:12
相关论文
共 26 条
[11]   ELISA-BASED DETERMINATION OF IMMUNOLOGICAL BINDING CONSTANTS [J].
LI, CKN .
MOLECULAR IMMUNOLOGY, 1985, 22 (03) :321-327
[12]   SURFACE-PLASMON RESONANCE FOR GAS-DETECTION AND BIOSENSING [J].
LIEDBERG, B ;
NYLANDER, C ;
LUNDSTROM, I .
SENSORS AND ACTUATORS, 1983, 4 (02) :299-304
[13]   A NOVEL HYDROGEL MATRIX ON GOLD SURFACES IN SURFACE-PLASMON RESONANCE SENSORS FOR FAST AND EFFICIENT COVALENT IMMOBILIZATION OF LIGANDS [J].
LOFAS, S ;
JOHNSSON, B .
JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1990, (21) :1526-1528
[14]   PROTEIN ADSORPTION ON CROSSLINKED POLYDIMETHYLSILOXANE USING TOTAL INTERNAL-REFLECTION FLUORESCENCE [J].
LOK, BK ;
CHENG, YL ;
ROBERTSON, CR .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1983, 91 (01) :104-116
[15]  
MASON DW, 1986, HDB EXPT IMMUNOLOGY, V1
[16]   THEORY OF STATIONARY CURRENT VOLTAGE CURVES OF REDOX ELECTRODE REACTIONS IN HYDRODYNAMIC VOLTAMMETRY .2. LAMINAR TUBE AND CHANNEL CURRENTS [J].
MATSUDA, H .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1967, 15 (04) :325-&
[17]   IMMUNOASSAY BASED ON SURFACE-PLASMON OSCILLATIONS [J].
MAYO, CS ;
HALLOCK, RB .
JOURNAL OF IMMUNOLOGICAL METHODS, 1989, 120 (01) :105-114
[18]   DISSOCIATION KINETICS OF ANTIGEN-ANTIBODY INTERACTIONS - STUDIES ON A PANEL OF ANTI-ALBUMIN MONOCLONAL-ANTIBODIES [J].
OLSON, WC ;
SPITZNAGEL, TM ;
YARMUSH, ML .
MOLECULAR IMMUNOLOGY, 1989, 26 (02) :129-136
[19]   A METHOD FOR THE DETERMINATION OF ANTIBODY-AFFINITY USING A DIRECT ELISA [J].
SCHOTS, A ;
VANDERLEEDE, BJ ;
DEJONGH, E ;
EGBERTS, E .
JOURNAL OF IMMUNOLOGICAL METHODS, 1988, 109 (02) :225-233
[20]  
SKUBITZ KM, 1975, J IMMUNOL, V114, P1369