RAPID-DETERMINATION OF ZINC AND IRON IN FOODS BY FLOW-INJECTION ANALYSIS WITH FLAME ATOMIC-ABSORPTION SPECTROPHOTOMETRY AND SLURRY NEBULIZATION

被引:33
作者
DEANDRADE, JC [1 ]
STRONG, FC [1 ]
MARTIN, NJ [1 ]
机构
[1] UNIV ESTADUAL CAMPINAS,FAC ENGN DEALIMENTOS,BR-13081 CAMPINAS,SP,BRAZIL
关键词
D O I
10.1016/0039-9140(90)80099-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A rapid method of determining zinc and iron in food by flame atomic-absorption spectrophotometry with slurry nebulization into an air-acetylene flame has been developed. A V-groove, clog-free Babington-type nebulizer, coupled to a single-line flow-injection analysis (FIA) system, was employed to introduce the slurry into the spray chamber. Under the FIA conditions described, an injection frequency of 120/hr is possible, with negligible carry-over and memory effects. The calibration graphs were obtained by using various concentrations (up to 0.1 g/ml) of white bean homogenate as standards, rather than solutions. The method has been applied to various kinds of foods, including grains, vegetables, fruits and sausage. Homogenization of semi-prepared samples to form slurries took only 4 min. Relative deviations between results by the slurry and solution methods for both elements averaged 2-3%. Detection limits by the slurry method were 0.3 μg/ml Zn and 0.6 μg/ml Fe. © 1990.
引用
收藏
页码:711 / 718
页数:8