CONTINUOUS AUTOMATIC-DETERMINATION OF AMMONIA BY USING AN INTEGRATED SEPARATION DETECTION UNIT

被引:2
作者
DELATORRE, M
DECASTRO, MDL
VALCARCEL, M
机构
[1] Department of Analytical Chemistry, Faculty of Sciences, University of Córdoba
关键词
D O I
10.1016/0039-9140(92)80108-P
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automatic continuous-flow photometric method for the determination of ammonia is proposed. It is based on the Berthelot reaction, the product of which is temporarily immobilized on a suitable support (Sephadex QAE) located in the flow-cell. The retained product is eluted after measurement by a cationic surfactant contained in the carrier solution. The method allows the determination of the analyte between 0.4 and 20.0-mu-g/ml, with an RSD of 0.8%, at a sampling frequency of 13/hr. The determination limit can be reduced by a factor of ten by using a flow-cell with a two-fold longer path length, but the sampling frequency is also reduced as a result. The method was applied to the determination of this analyte in agricultural samples (plants and soils) and compared with standard methods for these types of samples.
引用
收藏
页码:869 / 873
页数:5
相关论文
共 22 条
[21]   FIBEROPTIC FLUORESCING SENSOR FOR AMMONIA [J].
WOLFBEIS, OS ;
POSCH, HE .
ANALYTICA CHIMICA ACTA, 1986, 185 :321-327
[22]   FLOW-INJECTION ANALYSIS AS A DIAGNOSTIC-TECHNIQUE FOR DEVELOPMENT AND TESTING OF CHEMICAL SENSORS [J].
YERIAN, TD ;
CHRISTIAN, GD ;
RUZICKA, J .
ANALYTICA CHIMICA ACTA, 1988, 204 (1-2) :7-28