EVALUATION OF A POINT-OF-CARE TESTING PRODUCT FOR DRUGS OF ABUSE - TESTING SITE IS A KEY VARIABLE

被引:28
作者
KRANZLER, HR [1 ]
STONE, J [1 ]
MCLAUGHLIN, L [1 ]
机构
[1] UNIV CONNECTICUT,CTR HLTH,DEPT LAB MED,FARMINGTON,CT 06030
关键词
DRUGS OF ABUSE; SCREENING; IMMUNOASSAY; POINT-OF-CARE;
D O I
10.1016/0376-8716(95)01188-9
中图分类号
R194 [卫生标准、卫生检查、医药管理];
学科分类号
摘要
We evaluated the EZ-SCREEN Test Kit (Editek, Inc., Burlington, NC), a point-of-care screening test for the presence of cocaine metabolite (benzoylecgonine; BE), cannabinoids (tetrahydrocannabinol; THC) and opiates in urine. Patient specimens (n = 34) were tested in the laboratory using both Abbott Laboratories ADx (ADx) and EZ-SCREEN (EZS-LAB), and by the nursing staff of an inpatient substance abuse treatment program using the EZ-SCREEN (EZS-RN). We found comparable analytical efficiency between methods used in the laboratory (> 95% for all three analytes with EZS-LAB and ADx) but lower efficiency for THC and BE with point-of-care testing (approximate to 82% for EZS-RN). Efficiency for EZS-RN opiates was 100%. We conclude that the EZ-SCREEN Test Kit may not be suitable for use in a busy clinical setting, unless specific measures are taken to insure the accuracy of point-of-care testing (e.g., minimal interruption or distraction, careful training). We recommend that prior to routine use of point-of-care testing products they be evaluated under normal working conditions with personnel who will eventually be required to perform routine testing.
引用
收藏
页码:55 / 62
页数:8
相关论文
共 18 条
[1]  
BARR JT, 1991, ANCILLARY BEDSIDE BL, P1
[2]  
BENDER JL, 1991, CLIN LABORATORY PROC, P27
[3]   LABORATORY ASSESSMENT OF GLUCOSE METERS DOES NOT PREDICT RELIABILITY OF CLINICAL-PERFORMANCE [J].
BERNBAUM, M ;
ALBERT, SG ;
MCGINNIS, J ;
BRUSCA, S ;
MILLER, D ;
HOFFMANN, J ;
KAISER, FE ;
MOORADIAN, AD .
LABORATORY MEDICINE, 1994, 25 (01) :32-34
[4]  
BUDGETT WT, 1992, J FORENSIC SCI, V37, P632
[5]   FORENSIC DRUG-TESTING FOR OPIATES .5. URINE TESTING FOR HEROIN, MORPHINE, AND CODEINE WITH COMMERCIAL OPIATE IMMUNOASSAYS [J].
CONE, EJ ;
DICKERSON, S ;
PAUL, BD ;
MITCHELL, JM .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1993, 17 (03) :156-164
[6]  
CONE EJ, 1989, J FORENSIC SCI, V34, P32
[7]  
DAVIAUD J, 1993, CLIN CHEM, V39, P53
[8]  
INNANEN VT, 1989, CAN MED ASSOC J, V140, P899
[9]   VALIDITY TESTING OF THE EZ-SCREEN(R) CANNABINOID TEST [J].
JENKINS, AJ ;
MILLS, LC ;
DARWIN, WD ;
HUESTIS, MA ;
CONE, EJ ;
MITCHELL, JM .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1993, 17 (05) :292-298
[10]  
JONES B, 1991, Q PROBE BEDSIDE GLUC, P5