DESOLVATION EFFECT ON THE ANALYTICAL PERFORMANCE OF MICROWAVE-INDUCED PLASMA-ATOMIC ABSORPTION SPECTROMETRY (MIP-AAS)

被引:13
作者
DUAN, YX [1 ]
ZHANG, HQ [1 ]
HUO, MY [1 ]
JIN, QH [1 ]
机构
[1] JILIN UNIV,DEPT CHEM,CHANGCHUN 130023,PEOPLES R CHINA
关键词
D O I
10.1016/0584-8547(94)80051-0
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
This paper describes an improvement of detection capability for microwave-induced plasma atomic absorption spectrometry (MIP-AAS) by using a modified desolation system, which includes heated tubing, a water-cooled condenser and a concentrated sulfuric acid desiccator. The influence of heating temperature, carrier-gas flow rate and microwave power on the analytical performance of MIP-AAS has been examined and the relationship between these experimental parameters discussed. The desolation efficiencies of the different steps, such as in the water-cooling system and sulfuric acid desiccator, have been quantitatively measured using silica gel traps. Under the optimized experimental conditions, an overall desolation efficiency of as high as 99% is achieved by using this recommended desolation system. The analytical figures of merit for MIP-AAS are compared and evaluated under different desolation conditions.
引用
收藏
页码:583 / 592
页数:10
相关论文
共 28 条
[1]   DETERMINATION OF TRACES OF CHLORIDE IN SOLUTION BY MICROWAVE-INDUCED PLASMA EMISSION-SPECTROMETRY USING CHLORINE GENERATION [J].
ALDER, JF ;
JIN, QH ;
SNOOK, RD .
ANALYTICA CHIMICA ACTA, 1980, 120 (NOV) :147-154
[2]  
ALKEMADE CTJ, 1979, FUNDAMENTALS ANAL FL, pCH4
[3]  
BROWNER RF, 1982, ANAL CHEM, V54, P533
[4]  
BROWNER RF, 1987, INDUCTIVELY COUPLED, pCH8
[5]   HELIUM MICROWAVE INDUCED PLASMA MASS-SPECTROMETRY FOR DETECTION OF METALS AND NONMETALS IN AQUEOUS-SOLUTIONS [J].
CREED, JT ;
DAVIDSON, TM ;
SHEN, WL ;
BROWN, PG ;
CARUSO, JA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (09) :909-924
[6]   DESIGN CONSIDERATIONS AND PRELIMINARY CHARACTERIZATIONS OF A KILOWATT-PLUS MICROWAVE-INDUCED PLASMA [J].
CULL, KB ;
CARNAHAN, JW .
APPLIED SPECTROSCOPY, 1988, 42 (06) :1061-1065
[7]   ELEMENTAL ANALYSIS WITH A MICROWAVE-INDUCED PLASMA-QUADRUPOLE MASS-SPECTROMETER SYSTEM [J].
DOUGLAS, DJ ;
FRENCH, JB .
ANALYTICAL CHEMISTRY, 1981, 53 (01) :37-41
[8]   EVALUATION OF THE PERFORMANCE OF MICROWAVE-INDUCED PLASMA-ATOMIC ABSORPTION SPECTROMETRY (MIP-AAS) [J].
DUAN, YX ;
HOU, MY ;
DU, ZH ;
JIN, QH .
APPLIED SPECTROSCOPY, 1993, 47 (11) :1871-1879
[9]   EVALUATION OF A LOW-POWERED ARGON MICROWAVE PLASMA DISCHARGE AS AN ATOMIZER FOR THE DETERMINATION OF MERCURY BY ATOMIC FLUORESCENCE SPECTROMETRY [J].
DUAN, YX ;
KONG, XX ;
ZHANG, HQ ;
LIU, J ;
JIN, QH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (01) :7-10
[10]  
Fuchs N. A, 1964, MECH AEROSOLS