ARTIFACTS IN FETAL PULSE OXIMETRY - INCOMPLETE SENSOR-TO-SKIN CONTACT

被引:36
作者
GARDOSI, JO
DAMIANOU, D
SCHRAM, CMH
机构
[1] UNIV NOTTINGHAM,DEPT OBSTET & GYNAECOL,PERINATAL RES & MONITORING UNIT,NOTTINGHAM NG7 2RD,ENGLAND
[2] UNIV NOTTINGHAM,BIOMED ENGN UNIT,NOTTINGHAM NG7 2RD,ENGLAND
关键词
PULSE OXIMETRY; ARTIFACT; INTRAPARTUM MONITORING; REFLECTANCE SENSORS; FETAL PROBE;
D O I
10.1016/S0002-9378(94)70116-4
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
OBJECTIVE: Our purpose was to investigate apposition of reflectance sensors as a possible source of artifact in fetal pulse oximetry. STUDY DESIGN: A laboratory model was created to examine the effect of varying the distance of a reflectance probe from a subject's finger. The setup used two wavelengths (660 and 940 nm) and pulsatile (AC) and nonpulsatile signals (DC). An investigator's finger was stabilized in a purposely built mold from where the distance between skin and sensors could be varied in 1 mm steps. RESULTS: The AC and DC components of the signal react differently to increasing the distance between sensor and skin. The overall effect is that the ratio R = AC red/DC red/AC infrared/DC infrared increases with skin to sensor distance, which results in falsely low readings of oxygen saturation. CONCLUSIONS: Poor sensor-to-skin contact in reflectance pulse oximetry may cause false readings despite the presence of good waveforms and hence go unrecognized. Probe and software design needs to be able to identify and exclude such artifact before this technique can become useful for intrapartum fetal monitoring.
引用
收藏
页码:1169 / 1173
页数:5
相关论文
共 8 条
[1]  
Dildy G. A., 1993, American Journal of Obstetrics and Gynecology, V168, P340
[2]  
GARDOSI J, 1989, LANCET, V2, P692
[3]   ADAPTATION OF PULSE OXIMETRY FOR FETAL MONITORING DURING LABOR [J].
GARDOSI, JO ;
SCHRAM, CM ;
SYMONDS, EM .
LANCET, 1991, 337 (8752) :1265-1267
[4]   FETAL MONITORING WITH PULSE OXIMETRY [J].
JOHNSON, N ;
JOHNSON, VA ;
FISHER, J ;
JOBBINGS, B ;
BANNISTER, J ;
LILFORD, RJ .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1991, 98 (01) :36-41
[5]  
KELLEHER JF, 1989, ANESTHESIOLOGY, V51, P787
[6]   DO FETAL PULSE OXIMETRY READINGS AT DELIVERY CORRELATE WITH CORD BLOOD OXYGENATION AND ACIDEMIA [J].
MCNAMARA, H ;
CHUNG, DC ;
LILFORD, R ;
JOHNSON, N .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1992, 99 (09) :735-738
[7]  
MENDELSON Y, 1992, Biomedical Instrumentation and Technology, V26, P215
[8]  
PEAT S, 1988, LANCET, V2, P213