IMPROVED DETECTION LIMITS FOR ELECTROSPRAY-IONIZATION ON A MAGNETIC-SECTOR MASS-SPECTROMETER BY USING AN ARRAY DETECTOR

被引:14
作者
CODY, RB
TAMURA, J
FINCH, JW
MUSSELMAN, BD
机构
[1] JEOL USA, Inc., Peabody, 01960, MA
关键词
D O I
10.1016/1044-0305(94)85033-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Array detection was compared with point detection for solutions of hen egg-white lysozyme, equine myoglobin, and ubiquitin analyzed by electrospray ionization with a magnetic sector mass spectrometer. The detection limits for samples analyzed by using the array detector system were at least 10 times lower than could be achieved by using a point detector on the same mass spectrometer. The minimum detectable quantity of protein corresponded to a signal-to-background ratio of approximately 2:1 for a 500 amol/muL solution of hen egg-white lysozyme. However, the ultimate practical sample concentrations appeared to be in the 10-100 fmol/muL range for the analysis of dilute solutions of relatively pure proteins or simple mixtures.
引用
收藏
页码:194 / 200
页数:7
相关论文
共 35 条
[1]  
BATEMAN RH, 1992, 40TH P ANN ASMS C MA, P1669
[2]  
BERKEL GJV, 1990, ANAL CHEM, V62, P1284
[3]   METHOD FOR THE ELECTROSPRAY IONIZATION OF HIGHLY CONDUCTIVE AQUEOUS-SOLUTIONS [J].
CHOWDHURY, SK ;
CHAIT, BT .
ANALYTICAL CHEMISTRY, 1991, 63 (15) :1660-1664
[4]   ELECTROSPRAY IONIZATION MAGNETIC-SECTOR MASS-SPECTROMETRY - CALIBRATION, RESOLUTION, AND ACCURATE MASS MEASUREMENTS [J].
CODY, RB ;
TAMURA, J ;
MUSSELMAN, BD .
ANALYTICAL CHEMISTRY, 1992, 64 (14) :1561-1570
[5]  
CODY RB, 1992, 40TH P ANN ASMS C MA, P1679
[6]  
COTRELL JS, 1987, ANAL CHEM, V59, P1990
[7]  
Covey T R, 1988, Rapid Commun Mass Spectrom, V2, P249, DOI 10.1002/rcm.1290021111
[8]  
DING JM, 1992, 40TH P ASMS C MASS S, P1671
[9]  
DOBBERSTEIN P, 1992, 40TH P ANN ASMS C MA, P1683
[10]   MOLECULAR BEAMS OF MACROIONS [J].
DOLE, M ;
MACK, LL ;
HINES, RL .
JOURNAL OF CHEMICAL PHYSICS, 1968, 49 (05) :2240-&