CALIBRATION OF A SI(LI) DETECTOR SYSTEM WITH DIFFERENT RADIATION ENTRANCE WINDOWS

被引:8
作者
EGGERT, F [1 ]
MANECK, M [1 ]
SCHOLZE, F [1 ]
KRUMREY, M [1 ]
TEGELER, E [1 ]
机构
[1] PHYS TECH BUNDESANSTALT,INST BERLIN,W-1000 BERLIN 10,GERMANY
关键词
D O I
10.1063/1.1142077
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The spectral quantum detection efficiency of a Si(Li) detector was determined using the electron storage ring BESSY as a standard source of calculable spectral photon flux. The calibrations were performed for the same detector system with a Be window, an ultrathin silicon-based window and without window for photon energies above 300 eV. For comparison the transmission of the windows was measured in monochromatic radiation from 350-2000 eV with high spectral resolution.