ABSORBED DOSE PERTURBATION CAUSED BY DIODES FOR SMALL-FIELD PHOTON DOSIMETRY

被引:54
作者
BEDDAR, AS
MASON, DJ
OBRIEN, PF
机构
[1] PRINCESS MARGARET HOSP, ONTARIO CANC INST, DEPT CLIN PHYS, TORONTO M4X 1K9, ONTARIO, CANADA
[2] TORONTO BAYVIEW REG CANC CTR, DIV MED PHYS, TORONTO M4N 3M5, ON, CANADA
关键词
STEREOTAXIC RADIOSURGERY; SMALL FIELD PENUMBRA; DOSE PERTURBATION; DIODE DETECTORS; PHOTON BEAM DOSIMETRY;
D O I
10.1118/1.597350
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Uncertainties in dose measurements which may be considered acceptable for standard radiotherapy techniques may not be acceptable for radiosurgery or other precision radiotherapy techniques. One example of this type of uncertainty are the small perturbations to measured dose distributions which are caused by the use of diode detectors. In this article the effect of using diodes to measure small field penumbra is assessed and comments are made about diode orientation and diode construction.
引用
收藏
页码:1075 / 1079
页数:5
相关论文
共 13 条
[1]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .2. PROPERTIES AND MEASUREMENTS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1901-1913
[2]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1883-1900
[3]  
BEDDAR AS, 1992, MED PHYS, V19, P790
[4]   COMPARISON OF THE PENUMBRA BETWEEN FOCUSED AND NONDIVERGENT BLOCKS - IMPLICATIONS FOR MULTILEAF COLLIMATORS [J].
BIGGS, P ;
CAPALUCCI, J ;
RUSSELL, M .
MEDICAL PHYSICS, 1991, 18 (04) :753-758
[5]   ANALYSIS OF PHYSICAL PARAMETERS ASSOCIATED WITH THE MEASUREMENT OF HIGH-ENERGY X-RAY PENUMBRA [J].
DAWSON, DJ ;
HARPER, JM ;
AKINRADEWO, AC .
MEDICAL PHYSICS, 1984, 11 (04) :491-497
[6]   DOSIMETRY OF A DUAL PHOTON ENERGY LINEAR-ACCELERATOR [J].
FINDLEY, DO ;
FORELL, BW ;
WONG, PS .
MEDICAL PHYSICS, 1987, 14 (02) :270-273
[7]   CHARACTERISTICS OF THE 6-MV PHOTON-BEAM PRODUCED BY A DUAL ENERGY LINEAR-ACCELERATOR [J].
IKORO, NC ;
JOHNSON, DA ;
ANTICH, PP .
MEDICAL PHYSICS, 1987, 14 (01) :93-97
[8]   RADIOSURGERY WITH UNFLATTENED 6-MV PHOTON BEAMS [J].
OBRIEN, PF ;
GILLIES, BA ;
SCHWARTZ, M ;
YOUNG, C ;
DAVEY, P .
MEDICAL PHYSICS, 1991, 18 (03) :519-521
[9]   DOSE DISTRIBUTIONS IN DYNAMIC STEREOTAXIC RADIOSURGERY [J].
PIKE, B ;
PODGORSAK, EB ;
PETERS, TM ;
PLA, C .
MEDICAL PHYSICS, 1987, 14 (05) :780-789
[10]  
PLAZAS MC, 1988, AAPM ANN M SAN ANTON