DETERMINATION OF ACIDIC PESTICIDES IN WATER BY A BENCHTOP ELECTROSPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETER

被引:79
作者
CRESCENZI, C [1 ]
DICORCIA, A [1 ]
MARCHESE, S [1 ]
SAMPERI, R [1 ]
机构
[1] UNIV ROMA LA SAPIENZA,DIPARTIMENTO CHIM,I-00185 ROME,ITALY
关键词
D O I
10.1021/ac00109a010
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A very sensitive and specific analytical procedure for determining 20 acidic pesticides in aqueous environmental samples using pneumatically assisted electrospray (ES) LC/MS is presented. This procedure involves passing 1- and 4-L river water and drinking water samples, respectively, through a 1-g graphitized carbon black (GCB) extraction cartridge. By exploiting the presence of positively charged active centers on the GCB surface, isolation of the acidic pesticides from base/neutral species was made possible by differential elution. Recoveries of the analytes were higher than 85%, irrespective of the aqueous matrix in which they were dissolved. Adoption of the ion pair technique by addition of 0.1 mmol/L K2HPO4 and 0.2 mmol/L tetrabutylammonium fluoride to the mobile phase allowed analytes to be analyzed as pre-formed ions, A conventional 4.6-mm-i.d. reversed-phase LC C-18 column operating with a 1 mL/min now of the mobile phase was used for chromatographing the analytes. A now of 30 mu L/min of the column effluent was diverted to the ES source, while the rest of the mobile phase was delivered to a UV detector set at 220-nm wavelength, The effects of the nature of the ion pair forming agent and its counterion as well as its concentration in the mobile phase on the response of the ES/MS detector were investigated. The effects of varying the skimmer cone voltage on the production of diagnostic fragments and the response of the MS detector were also evaluated. For the analytes considered, the response of the mass detector was linearly related to the amount of the analytes injected between 2.5 and 200 ng. A certain variation of the ion signal for the analytes considered occurred after several hours of continuous use of the LC/ES/MS instrumentation. For routine use, then, analyte quantitation could be better performed by the UV trace method while entrusting unambiguous identicalness of the analytes to the MS detector. The limits of sensitivity (signal-to-noise ratio = 3) of the method for the pesticides considered in drinking water and surface water samples were estimated to be about 2-6 and 8-25 ng/L, respectively.
引用
收藏
页码:1968 / 1975
页数:8
相关论文
共 14 条
  • [1] THE NATURE OF SURFACE CHEMICAL HETEROGENEITIES OF GRAPHITIZED CARBON-BLACK
    CAMPANELLA, L
    DICORCIA, A
    SAMPERI, R
    GAMBACORTA, A
    [J]. MATERIALS CHEMISTRY, 1982, 7 (03): : 429 - 438
  • [2] ION CHROMATOGRAPHY MASS-SPECTROMETRY FOR THE DETERMINATION OF ORGANIC AMMONIUM AND SULFATE COMPOUNDS
    CONBOY, JJ
    HENION, JD
    MARTIN, MW
    ZWEIGENBAUM, JA
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (08) : 800 - 807
  • [3] DICORCIA A, 1994, J AOAC INT, V77, P446
  • [4] EVALUATION OF GRAPHITIZED CARBON-BLACK AS A SELECTIVE ADSORBENT FOR EXTRACTING ACIDIC ORGANIC-COMPOUNDS FROM WATER
    DICORCIA, A
    MARCHESE, S
    SAMPERI, R
    [J]. JOURNAL OF CHROMATOGRAPHY, 1993, 642 (1-2): : 163 - 174
  • [5] MULTIRESIDUE METHOD FOR PESTICIDES IN DRINKING-WATER USING A GRAPHITIZED CARBON-BLACK CARTRIDGE EXTRACTION AND LIQUID-CHROMATOGRAPHIC ANALYSIS
    DICORCIA, A
    MARCHETTI, M
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (06) : 580 - 585
  • [6] GRAPHITIZED CARBON-BLACK EXTRACTION CARTRIDGES FOR MONITORING POLAR PESTICIDES IN WATER
    DICORCIA, A
    SAMPERI, R
    MARCOMINI, A
    STELLUTO, S
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (07) : 907 - 912
  • [7] METHOD DEVELOPMENT FOR MONITORING PESTICIDES IN ENVIRONMENTAL WATERS - LIQUID SOLID EXTRACTION FOLLOWED BY LIQUID-CHROMATOGRAPHY
    DICORCIA, A
    MARCHETTI, M
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1992, 26 (01) : 66 - 74
  • [8] ATMOSPHERIC-PRESSURE ION-SAMPLING SYSTEM FOR LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY ANALYSES ON A BENCHTOP MASS-SPECTROMETER
    DUFFIN, KL
    WACHS, T
    HENION, JD
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (01) : 61 - 68
  • [9] SPECIES-SELECTIVITY EFFECTS IN THE PRODUCTION OF ELECTROSPRAY IONS
    HIRAOKA, K
    MURATA, K
    KUDAKA, I
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1993, 7 (05) : 363 - 373
  • [10] HIGH-FLOW ION SPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY
    HOPFGARTNER, G
    WACHS, T
    BEAN, K
    HENION, J
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (04) : 439 - 446