A METHOD FOR THE UNAMBIGUOUS IDENTIFICATION OF TRYPTOPHAN IN AUTOMATED PROTEIN-SEQUENCE ANALYSIS

被引:2
作者
MULLERMICHEL, T [1 ]
BOHLEN, P [1 ]
机构
[1] AMER CYANAMID CO,DIV MED RES,BLDG 205,PEARL RIVER,NY 10965
关键词
D O I
10.1016/0003-2697(90)90404-W
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Some widely used standard protocols for the separation of phenylthiohydantoin amino acid derivatives by reverse-phase gradient HPLC do not provide separation of the phenylthiohydantoin derivative of tryptophan (PTH-Trp) from diphenylurea (DPU), a by-product generated during Edman degradation of proteins in variable amounts. Furthermore, PTH-Trp is usually recovered in low yield under typical experimental conditions used with automated sequencing equipment. These factors may compromise the unambiguous assignment of tryptophan residues in automated protein sequence analysis, especially when sequencing is performed at high sensitivity. We devised a reverse-phase HPLC method which allows the separation of DPU and PTH-Trp and therefore the correct assignment of PTH-Trp. The method is based on a modification of the HPLC gradient used to elute and separate all PTH amino acids of interest. With Applied Biosystems Model 477A protein sequencers with on-line PTH amino acid identification, the correct assignment of tryptophan was consistent and reproducible even when sequencing at very high sensitivity (5 pmol). © 1990.
引用
收藏
页码:169 / 173
页数:5
相关论文
共 16 条
[11]   ANALYSIS OF PHENYLTHIOHYDANTOIN AMINO-ACIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ON DUPONT ZORBAX CYANOPROPYLSILANE COLUMNS [J].
JOHNSON, ND ;
HUNKAPILLER, MW ;
HOOD, LE .
ANALYTICAL BIOCHEMISTRY, 1979, 100 (02) :335-338
[12]   IDENTIFICATION OF THE PHENYLTHIOHYDANTOIN DERIVATIVES OF AMINO-ACIDS BY HIGH-PRESSURE LIQUID-CHROMATOGRAPHY, USING A TERNARY, ISOCRATIC SOLVENT SYSTEM [J].
LOTTSPEICH, F .
HOPPE-SEYLERS ZEITSCHRIFT FUR PHYSIOLOGISCHE CHEMIE, 1980, 361 (12) :1829-1834
[14]   EXAMINATION OF AUTOMATED POLYPEPTIDE SEQUENCING USING STANDARD PHENYL ISOTHIOCYANATE REAGENT AND SUBPICOMOLE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS [J].
TEMPST, P ;
RIVIERE, L .
ANALYTICAL BIOCHEMISTRY, 1989, 183 (02) :290-300
[15]  
Wittmann-Liebold Brigitte, 1982, METHODS PROTEIN SEQU, P27
[16]   RAPID ANALYSIS OF AMINO-ACID PHENYLTHIOHYDANTOINS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
ZIMMERMAN, CL ;
APPELLA, E ;
PISANO, JJ .
ANALYTICAL BIOCHEMISTRY, 1977, 77 (02) :569-573