APPLICATION OF A NAFION-MODIFIED CARBON-PASTE ELECTRODE FOR THE ADSORPTIVE STRIPPING VOLTAMMETRIC DETERMINATION OF FENOTEROL IN PHARMACEUTICAL PREPARATIONS AND BIOLOGICAL-FLUIDS

被引:17
作者
BOYD, D
RODRIGUEZ, JRB
BLANCO, PT
SMYTH, MR
机构
[1] UNIV OVIEDO,DEPT PHYS & ANALYT CHEM,E-33006 OVIEDO,SPAIN
[2] DUBLIN CITY UNIV,SCH CHEM SCI,DUBLIN 9,IRELAND
关键词
FENOTEROL; NAFION-MODIFIED CARBON PASTE ELECTRODE; COMPLEX MATRICES;
D O I
10.1016/0731-7085(94)00055-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The preconcentration of fenoterol on a Nafion-modified carbon paste electrode and its subsequent determination using differential pulse voltammetry is described. The effect of pH and percentage Nafion concentration on the accumulation behaviour of fenoterol was studied, and accumulation curves, calibration graphs and reproducibility studies al two different Nafion concentrations have been carried out in the range 2.5 x 10(-8)-5.0 x 10(-7) M fenoterol. A limit of detection in aqueous solutions, calculated using a signal-to-noise ratio (SM) of 3, was 9.0 x 10(-9) M. Application of the electrode to pharmaceutical preparations, without sample pretreatment, resulted in acceptable deviation from the stated concentration (RSD = +/- 3.81%, n = 4). For more complex matrices, a suitable extraction procedure was developed, resulting in recoveries of >90% (urine) and >75% (serum).
引用
收藏
页码:1069 / 1074
页数:6
相关论文
共 20 条
[1]   COMPARISON OF ISOTACHOPHORESIS, CAPILLARY ZONE ELECTROPHORESIS AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FOR THE DETERMINATION OF SALBUTAMOL, TERBUTALINE SULFATE AND FENOTEROL HYDROBROMIDE IN PHARMACEUTICAL DOSAGE FORMS [J].
ACKERMANS, MT ;
BECKERS, JL ;
EVERAERTS, FM ;
SEELEN, IGJA .
JOURNAL OF CHROMATOGRAPHY, 1992, 590 (02) :341-353
[2]   ELECTROCHEMISTRY OF HYDROPHOBIC NAFION GELS .3. PRECONCENTRATION AND ELECTROCHEMICAL DETECTION OF CATIONIC AND NEUTRAL ORGANIC-COMPOUNDS IN HYDROPHOBIC NAFION GELS MADE FROM VARIOUS NORMAL-ALKYLPHOSPHATES [J].
AUDEBERT, P ;
DIVISIABLOHORN, B ;
ALDEBERT, P ;
MICHALAK, F .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1992, 322 (1-2) :301-309
[3]  
BOENISCH B, 1992, P FLAIR WORKSH THESS, P102
[4]   APPLICATION OF MATRIX SOLID-PHASE DISPERSION FOR THE DETERMINATION OF CLENBUTEROL IN LIVER SAMPLES [J].
BOYD, D ;
SHEARAN, P ;
HOPKINS, JP ;
OKEEFFE, M ;
SMYTH, MR .
ANALYTICA CHIMICA ACTA, 1993, 275 (1-2) :221-226
[5]  
BOYD D, IN PRESS ANALYST
[6]   NAFION-COATED CARBON-FIBER ELECTRODES FOR NEUROCHEMICAL STUDIES IN BRAIN-TISSUE [J].
CAPELLA, P ;
GHASEMZADEH, B ;
MITCHELL, K ;
ADAMS, RN .
ELECTROANALYSIS, 1990, 2 (03) :175-182
[7]  
Gao Z., 1990, ANAL CHIM ACTA, V241, P137
[8]   APPLICATION OF POLYMER-COATED GLASSY-CARBON ELECTRODES IN ANODIC-STRIPPING VOLTAMMETRY [J].
HOYER, B ;
FLORENCE, TM ;
BATLEY, GE .
ANALYTICAL CHEMISTRY, 1987, 59 (13) :1608-1614
[9]  
HUILIANG H, 1988, ANAL CHIM ACTA, V207, P17
[10]   ELECTRODEPOSITION OF PT ULTRAMICROPARTICLES IN NAFION FILMS ON GLASSY-CARBON ELECTRODES [J].
ITAYA, K ;
TAKAHASHI, H ;
UCHIDA, I .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1986, 208 (02) :373-382